Improved porous silicon (P-Si) microarray based PSA (prostate specific antigen) immunoassay by optimized surface density of the capture antibody

机译:通过优化捕获抗体的表面密度改进了基于多孔硅(P-Si)微阵列的PSA(前列腺特异性抗原)免疫测定

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Enriching the surface density of immobilized capture antibodies enhances the detection signal of antibody sandwich microarrays. In this study, we improved the detection sensitivity of our previously developed P-Si (porous silicon) antibody microarray by optimizing concentrations of the capturing antibody. We investigated immunoassays using a P-Si microarray at three different capture antibody (PSA - prostate specific antigen) concentrations, analyzing the influence of the antibody density on the assay detection sensitivity. The LOD (limit of detection) for PSA was 2.5ngmL−1, 80pgmL−1, and 800fgmL−1 when arraying the PSA antibody, H117 at the concentration 15µgmL−1, 35µgmL−1 and 154µgmL−1, respectively. We further investigated PSA spiked into human female serum in the range of 800fgmL−1 to 500ngmL−1. The microarray showed a LOD of 800fgmL−1 and a dynamic range of 800 fgmL−1 to 80ngmL−1 in serum spiked samples.
机译:丰富固定的捕获抗体的表面密度可增强抗体夹心微阵列的检测信号。在这项研究中,我们通过优化捕获抗体的浓度来提高我们先前开发的P-Si(多孔硅)抗体微阵列的检测灵敏度。我们研究了在三种不同的捕获抗体(PSA-前列腺特异抗原)浓度下使用P-Si微阵列进行的免疫测定,分析了抗体密度对测定检测灵敏度的影响。排列PSA抗体时,PSA的LOD(检出限)为2.5ngmL -1 ,80pgmL -1 和800fgmL -1 , H117的浓度分别为15µgmL -1 ,35µgmL -1 和154µgmL -1 。我们进一步研究了掺入人女性血清的PSA范围为800fgmL -1 至500ngmL -1 。微阵列显示血清加标样品中的LOD为800fgmL -1 ,动态范围为800 fgmL -1 至80ngmL -1

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号