摘要

A 41Ca interlaboratory comparison between Lawrence Livermore National Laboratory (LLNL) and the Purdue Rare Isotope Laboratory (PRIME Lab) has been completed. Analysis of the ratios assayed by accelerator mass spectrometry (AMS) shows that there is no statistically significant difference in the ratios. Further, Bayesian analysis shows that the uncertainties reported by both facilities are correct with the possibility of a slight under-estimation by one laboratory. Finally, the chemistry procedures used by the two facilities to produce CaF2 for the cesium sputter ion source are robust and don't yield any significant differences in the final result.
机译:劳伦斯·利弗莫尔国家实验室(LLNL)和普渡大学稀有同位素实验室(PRIME Lab)之间的 41 Ca实验室间比较已完成。通过加速器质谱(AMS)测定的比率分析表明,比率没有统计学上的显着差异。此外,贝叶斯分析表明,两种设施报告的不确定性都是正确的,而且一个实验室可能会略低估量。最后,这两种设备用于生产铯溅射离子源的CaF2的化学程序是可靠的,并且在最终结果中不会产生任何显着差异。