AXIAL TRANSMISSION METHOD FOR LONG BONE FRACTURE EVALUATION BY ULTRASONIC GUIDED WAVES: SIMULATION PHANTOM AND IN VITRO EXPERIMENTS

机译:超声引导波评估长骨骨折的轴向传输方法:模拟幻影和体外实验

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Mode conversion occurs when the ultrasonic guided waves encounter fractures. The aim of this study was to investigate the feasibility of fracture assessment in long cortical bone using guided-mode conversion. Mode conversion behavior between the fundamental modes S0 and A0 was analyzed. The expressions proposed for modal velocity were used to identify the original and converted modes. Simulations and phantom experiments were performed using 1.0-mm-thick steel plates with a notch width of 0.5 mm and notch depths of 0.2, 0.4, 0.6 and 0.8 mm. Furthermore, in vitro experiments were carried out on nine ovine tibias with 1.0-mm-wide partial transverse gap break and cortical thickness varying from 2.10 to 3.88 mm. The study confirmed that mode conversion gradually becomes observable as fracture depth increases. Energy percentages of the converted modes correlated strongly with fracture depth, as illustrated by the frequency-sweeping experiments on steel phantoms (100–1100 kHz, r2 = 0.97, p < 0.0069) and the fixed-frequency experiments on nine ovine tibias (250 kHz, r2 = 0.97, p < 0.0056). The approaches described, including mode excitation, velocity expressions and energy percentage criteria, may also contribute to ultrasonic monitoring of long bone fracture healing.
机译:当超声导波遇到破裂时,会发生模式转换。这项研究的目的是研究使用引导模式转换的长皮质骨骨折评估的可行性。分析了基本模式S0和A0之间的模式转换行为。为模态速度提出的表达式用于识别原始模式和转换模式。使用厚度为1.0毫米,切口宽度为0.5毫米,切口深度为0.2、0.4、0.6和0.8毫米的钢板进行模拟和幻影实验。此外,还对九个绵羊胫骨进行了体外实验,这些胫骨的宽度为1.0毫米,部分横向间隙断裂且皮层厚度为2.10至3.88毫米。研究证实,随着裂缝深度的增加,模式转换逐渐变得可观察到。转换模态的能量百分比与断裂深度密切相关,如钢模(100-1100 kHz,r 2 = 0.97,p <0.0069)上的扫频实验所示在九个绵羊胫骨上进行实验(250 kHz,r 2 = 0.97,p <0.0056)。所描述的方法,包括模式激励,速度表达式和能量百分比标准,也可能有助于长骨骨折愈合的超声监测。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号