Repeatability Reproducibility and Comparability of Subjective and Objective Measurements of Intraocular Forward Scattering in Healthy Subjects

机译:健康受试者眼内前向散射的主观和客观测量的可重复性可再现性和可比性

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Purpose. To assess the repeatability, reproducibility, and comparability of measurements of subjective and objective forward scattering in healthy subjects. Methods. We prospectively examined twenty eyes of 20 healthy volunteers (7 men and 13 women; ages, 28.4 ± 4.1 years). The logarithmic straylight value (log(s)) and the objective scattering index (OSI) were measured with a straylight meter (C-Quant) and a point-spread function meter (OQAS), respectively. Results. The 95% limits of agreement (LoA) between first and second measurements ranged from −0.211 to 0.207 for the C-Quant and from −0.302 to 0.477 for the OQAS. The intraclass correlation coefficients for the repeatability of the log(s) and OSI measurements were 0.815 and 0.926, respectively. The mean difference between examiners was −0.051 ± 0.133 (95% LoA; −0.311 to 0.209) for the C-Quant and 0.080 ± 0.307 (−0.522 to 0.682) for the OQAS. There was a modest, but significant, correlation between the log(s) and the OSI (Spearman correlation coefficient r = 0.498, p = 0.026). Conclusions. The C-Quant and the OQAS provide good repeatability and reproducibility, although the OQAS measurement provides a slightly higher ICC than the C-Quant measurement. The subjective forward scattering may be to some extent expressed in the objective forward scattering in healthy subjects.
机译:目的。评估健康受试者中主观和客观前向散射测量值的可重复性,再现性和可比性。方法。我们前瞻性地检查了20名健康志愿者的20只眼睛(7名男性和13名女性;年龄28.4±4.1岁)。对数杂散光值(log)和目标散射指数(OSI)分别用杂散光计(C-Quant)和点扩展函数计(OQAS)测量。结果。第一次和第二次测量之间的95%一致性极限(LoA)对于C-Quant范围从-0.211到0.207,对于OQAS从-0.302到0.477。对数和OSI测量的可重复性的类内相关系数分别为0.815和0.926。对于C-Quant,检查者之间的平均差异为-0.051±0.133(95%LoA; -0.311至0.209),对于OQAS,其平均差异为0.080±0.307(-0.522至0.682)。 log和OSI之间存在适度但重要的相关性(Spearman相关系数r = 0.498,p = 0.026)。结论。尽管OQAS测量提供的ICC略高于C-Quant测量,但C-Quant和OQAS提供了良好的可重复性和可重复性。在健康受试者中,主观正向散射可以在某种程度上表示为客观正向散射。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号