This work investigates the efficacy of “coplanar shielding,” in which copper shields are oriented concentric and coplanar to the RF coils rather than implemented as a full ground plane behind them. Following FDTD simulations to determine optimal shielding parameters, two coil geometries were constructed: a circular loop surface coil and a half-volume five-element receive array. Each was evaluated using bench measurements with and without coplanar shielding. Imaging, including accelerated SENSE imaging, was performed with the shielded and unshielded receive arrays on a whole-body 7T scanner. Results from modeled and fabricated coils showed good agreement with improvements in Q factors for all cases. Imaging showed substantial improvements in SNR and g-factors for the coplanar shielded array.
机译:这项工作研究了“共面屏蔽”的功效,其中铜屏蔽与RF线圈同心且共面,而不是在它们后面实现完整的接地层。在进行FDTD仿真以确定最佳屏蔽参数之后,构建了两种线圈几何形状:圆形环形表面线圈和半体积五元素接收阵列。使用带有和不带有共面屏蔽的工作台测量评估每个对象。在全身7T扫描仪上使用屏蔽和非屏蔽接收阵列执行成像,包括加速的SENSE成像。在所有情况下,建模和制造线圈的结果均显示出与Q因子改善的良好一致性。成像显示共面屏蔽阵列的SNR和g因子显着提高。