The development of four‐dimensional (4D) imaging, treatment planning, and treatment delivery methods for radiation therapy requires a quality assurance device that can reproduce clinical motions. Here, we present a high‐precision translation stage for testing and validating 4D techniques. These techniques may require spatial resolutions of 1 mm and temporal resolutions of 2 – 30 Hz for computed tomography imaging, electromagnetic tracking, and fluoroscopic imaging. A one‐dimensional programmable translation stage capable of reproducing idealized and measured anatomic motions common to the thorax and providing support for phantoms weighing up to 27 kg was designed and built to meet the foregoing spatial and temporal resolution requirements. The stage consists of a polycarbonate base and table, driven by an ac servo motor with encoder feedback by means of a belt‐coupled precision screw. Complex motions are made possible by a programmable motion controller that can run multiple independent control and monitoring programs concurrently on independent motion axes. Programmable input and output ports allow motion to be synchronized with beam delivery and other imaging and treatment delivery devices to within 2.0 ms. Average deviations from the programmed positions are typically 0.2 mm or less, and the average maximum positional errors are typically 0.5 mm for an indefinite number of parameterized breathing motion cycles during reproduction of measured target volume motions for several minutes.PACS number: 87.66.Xa
机译:放射疗法的四维(4D)成像,治疗计划和治疗提供方法的开发需要质量保证设备,该设备可以再现临床动作。在这里,我们为测试和验证4D技术提供了一个高精度的翻译阶段。这些技术可能需要1 mm的空间分辨率和2 – 30 Hz的时间分辨率,以进行计算机断层扫描成像,电磁跟踪和荧光透视成像。设计并制造了一种一维可编程平移台,该台能够再现胸腔常见的理想化的和测量的解剖运动并支持重达27 kg的体模,以满足上述空间和时间分辨率的要求。该平台由聚碳酸酯底座和工作台组成,由交流伺服电机驱动,并通过皮带耦合的精密螺钉驱动编码器反馈。可编程运动控制器使复杂的运动成为可能,该运动控制器可以在独立的运动轴上同时运行多个独立的控制和监视程序。可编程的输入和输出端口允许运动与光束传输以及其他成像和治疗传输设备在2.0毫秒内同步。与编程位置的平均偏差通常为0.2 mm或更小,并且在重复测量的目标体积运动数分钟期间,无限次数的参数化呼吸运动周期的平均最大位置误差通常为0.5 mm.PACS编号:87.66.Xa

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号