2.5'/> Characterization of linear accelerator X‐ray source size using a laminated beam‐spot camera
首页> 美国卫生研究院文献>Journal of Applied Clinical Medical Physics >Characterization of linear accelerator X‐ray source size using a laminated beam‐spot camera
【2h】

Characterization of linear accelerator X‐ray source size using a laminated beam‐spot camera

机译:使用叠层束斑相机表征线性加速器X射线源的尺寸

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A laminated beam‐spot camera of length 20 cm and effective cross‐sectional area 2.5cm×3cm was designed and constructed for the measurement of X‐ray beam‐spot sizes on different models of Siemens accelerators. With the accelerator gantry at 180° and camera positioned on an accessory tray holder, an XV film placed in contact with the camera at the distal end of it detected those X‐rays that were transmitted through the camera. The FWHM of the detected X‐ray intensity profile in the gun–target (G–T) direction or the orthogonal A–B direction was used as a measure of the beam‐spot size in that direction. Siemens Mevatron MXEs exhibited a beam‐spot size of 1.7±0.2mm in both the in‐plane and cross‐plane directions for 6 MV photon beams. The beam‐spot size observed for a Mevatron MDX‐2 was larger by up to 1 mm, and also was different for the in‐plane and cross‐plane directions. For Siemens PRIMUS accelerators, the beam‐spot size in the in‐plane direction was found to fall in the range 2.02.2±0.2mm, whereas the beam‐spot size in the cross‐plane direction fell within 1.71.9±0.2mm for 6, 10, and 18 MV photon beams. Assessment of long‐term stability of the beam‐spot size shows the spot size remains fairly stable over time.PACS number: 87.56.B‐, 87.56.bd
机译:长度为20厘米且有效横截面积为 2.5 cm × 3 cm 是为在各种型号的西门子加速器上测量X射线束斑尺寸而设计和构造的。将加速器机架置于180°,并将照相机放置在附件托盘支架上,将XV胶片放在与照相机接触的远端,该XV胶片会检测到通过照相机传输的X射线。在枪目标(G–T)方向或正交A–B方向上检测到的X射线强度轮廓的FWHM被用作该方向上束斑尺寸的度量。西门子Mevatron MXE的光斑大小为1.7 < mo>± 0.2 mm 6 MV光子束的平面方向。 Mevatron MDX-2观察到的束斑大小最大增加了1 mm,并且在平面内和横断面方向上也不同。对于Siemens PRIMUS加速器,发现面内方向上的束斑大小在 2.0 2.2 ± 0.2 mm ,而跨平面方向的束斑大小在 1.7 - 1.9 ± 0.2 mm 用于6、10和18 MV光子束。评估束斑尺寸的长期稳定性表明,束斑尺寸随时间保持相当稳定.PACS编号:87.56.B-,87.56.bd

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号