首页> 美国卫生研究院文献>Journal of Applied Clinical Medical Physics >Commissioning and implementation of an implantable dosimeter for radiation therapy
【2h】

Commissioning and implementation of an implantable dosimeter for radiation therapy

机译:用于放射治疗的可植入剂量计的调试和实施

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

In this article we describe commissioning and implementation procedures for the Dose Verification System (DVS) with permanently implanted in vivo wireless, telemetric radiation dosimeters for absolute dose measurements. The dosimeter uses a semiconductor device called a metal–oxide semiconductor field‐effect transistor (MOSFET) to measure radiation dose. A MOSFET is a transistor that is generally used for amplifying or switching electronic signals. The implantable dosimeter was implemented with the goal of verifying the dose delivered to radiation therapy patients. For the purpose of acceptance testing, commissioning, and clinical implementation and to evaluate characteristics of the dosimeter, the following tests were performed: 1) temperature dependence, 2) reproducibility, 3) field size dependence, 4) postirradiation signal drift, 5) dependence on average dose rate, 6) linearity test, 7) angular dependence (different gantry angle position), 8) angular dependence (different DVS angle position), 9) dose rate dependence, 10) irradiation depth dependence, 11) effect of cone‐beam exposure to the dosimeter, and 12) multiple reading effect. The dosimeter is not currently calibrated for use in the kV range; nonetheless, the effect of the cone‐beam procedure on the MOSFET dosimeter was investigated. Phantom studies were performed in both air and water using an Elekta Synergy S Beam‐Modulator linear accelerator. Commissioning and clinical implementation for prostate cancer patients receiving external‐beam radiation therapy were performed in compliance with the general recommendations given for in vivo dosimetry devices. The reproducibility test in water at human body temperature (37°C) showed a 1.4% absolute difference, with a standard deviation of 5.72 cGy (i.e., SD = 2.9%). The constancy test shows that the average readings at room temperature were 3% lower compared to the readings at human body temperature, with a SD = 2%. Measurements were not dependent upon field size. Due to postirradiation signal drift, the following corrections are suggested: −2.8%, −2%, 0.5%, and 2.5% for the readings taken after 0.5, 1, 5, or 10 min, respectively. Different gantry angles did not influence the readings. The maximum error was less than 1% with a maximum SD = 3.61cGy (1.8%) for the gantry angle of 45°. However, readings are dependent on the dosimeter orientation. The average dose reading was 7.89 cGy (SD = 1.46cGy) when CBCT imaging was used for the pelvis protocol, and when postirradiation measurement was taken at 2.5 min (expected 2–3 cGy). The clinical implementation of the implantable MOSFET dosimeters for prostate cancer radiation therapy is described. Measurements performed for commissioning show that the dosimeter, if used within specifications, provides sufficient accuracy for its intended use in clinical procedures. The postradiation signal drift, temperature dependence, variation of reproducibility, and rotational isotropy could be encountered if the dosimeter is used outside the manufacturer's specifications. The dosimeter can be used as a tool for quantifying dose at depth, as well as to evaluate adherence between planned doses and the delivered doses. Currently, the system is clinically implemented with ±7% tolerance.PACS numbers: 87.53.‐j; 87.55.‐x
机译:在本文中,我们描述了剂量验证系统(DVS)的调试和实施程序,该系统带有永久植入的体内无线遥测辐射剂量计,用于绝对剂量测量。剂量计使用一种称为金属氧化物半导体场效应晶体管(MOSFET)的半导体器件来测量辐射剂量。 MOSFET是通常用于放大或切换电子信号的晶体管。实施植入式剂量计的目的是验证传递给放射治疗患者的剂量。为了进行验收测试,调试和临床实施以及评估剂量计的特性,进行了以下测试:1)温度依赖性,2)可再现性,3)场大小依赖性,4)照射后信号漂移,5)依赖性在平均剂量率上,6)线性测试,7)角度依赖性(不同机架角度位置),8)角度依赖性(不同DVS角度位置),9)剂量率依赖性,10)照射深度依赖性,11)锥度效应光束暴露于剂量计,以及12)多重读取效果。剂量计当前未在kV范围内使用进行校准;尽管如此,仍研究了锥束程序对MOSFET剂量计的影响。使用Elekta Synergy S Beam-Modulator线性加速器在空气和水中进行幻影研究。接受体外放射治疗的前列腺癌患者的调试和临床实施符合体内剂量测定仪的一般建议。在人体温度(37°C)的水中进行的重现性测试显示绝对差异为1.4%,标准偏差为5.72 cGy(即SD = 2.9%)。稳定性测试表明,室温下的平均读数比人体温度下的平均读数低3%,SD = 2%。测量结果不取决于字段大小。由于辐照后信号的漂移,建议进行以下校正:对于0.5、1、5或10分钟后的读数,分别为-2.8%,-2%,0.5%和2.5%。不同的机架角度不会影响读数。对于45°的龙门角度,最大误差小于1%,最大SD = 3.61cGy(1.8%)。但是,读数取决于剂量计的方向。当骨盆方案使用CBCT成像时,以及在2.5分钟进行辐射后测量时(预期2-3 cGy),平均剂量读数为7.89 cGy(SD = 1.46cGy)。描述了用于前列腺癌放射治疗的可植入MOSFET剂量计的临床实施方式。进行调试的测量结果表明,剂量计(如果在规格范围内使用)可为其在临床程序中的预期用途提供足够的准确性。如果剂量计的使用超出制造商的规格,则可能会遇到辐射后信号漂移,温度依赖性,再现性变化和旋转各向同性的问题。剂量计可以用作定量深处剂量以及评估计划剂量和所输送剂量之间依从性的工具。目前,该系统已在临床上实现了±7%的公差。PACS编号:87.53.-j; 87.55.-x

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号