首页> 美国卫生研究院文献>Journal of Applied Clinical Medical Physics >Evaluation of image lag in a flat‐panel detector‐equipped cardiovascular X‐ray machine using a newly developed dynamic phantom
【2h】

Evaluation of image lag in a flat‐panel detector‐equipped cardiovascular X‐ray machine using a newly developed dynamic phantom

机译:使用新开发的动态体模在配有探测器的平板X射线机中评估图像滞后

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

We developed a dynamic phantom for use in routine checks. This phantom can be used to physically evaluate image lag that occurs in dynamic images. It has a unique measurement method. ROIs on the target are chosen, and, with the position of ROIs fixed on the image, changes in pixel value are detected physically when the target passes through the ROIs over time and perceived as image lag. Thus, it was possible to physically detect different intensities of image lag lasting less than one second while maintaining the same intensity trends. The checking technique we propose with the dynamic phantom that we developed could be effective for routine checking of fluoroscopy X‐ray machines, and could become an established method.PACS number: 87.59.C‐, 87.59.Dj
机译:我们开发了一种动态体模,用于例行检查。该体模可用于物理评估动态图像中出现的图像滞后。它具有独特的测量方法。选择目标上的ROI,并在图像上固定ROI的位置的情况下,当目标随时间经过ROI并被感知为图像滞后时,会物理检测像素值的变化。因此,可以在保持相同的强度趋势的同时,物理地检测持续不到一秒的图像延迟的不同强度。我们提出的带有动态体模的检查技术可以有效地对X射线透视机进行常规检查,并且可以成为一种已建立的方法.PACS编号:87.59.C-,87.59.Dj