Accuracy of automatic matching of Catphan 504 phantom in cone‐beam computed tomography for tube current–exposure time product

机译:锥形束计算机体层摄影中的Catphan 504体模与管电流-曝光时间乘积的自动匹配精度

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The purpose of this study is to evaluate the accuracy of automatic matching in cone‐beam computed tomography (CBCT) images relative to the reduction of total tube current–exposure time product (mAs) for the X‐ray imaging (XI) system. The CBCT images were acquired with the Catphan 504 phantom various total mAs ratios such as 1.00, 0.83, 0.67, 0.57, and 0.50. For studying the automatic matching accuracy, the phantom images were acquired with a six‐dimensional shifting table. The image quality and correction of automatic matching were compared. With a decreasing total mAs ratio, the noise of the images increased and the low‐contrast resolution decreased, while the accuracy of the automatic matching did not change. Therefore, this study shows that a change of the total mAs while acquiring CBCT images has no effect on the automatic matching of Catphan 504 phantom in XI system.PACS number(s): 87.57.C‐, 87.57.Q‐
机译:这项研究的目的是评估与X射线成像(XI)系统的总管电流-曝光时间乘积(mAs)的减少有关的锥形束计算机断层扫描(CBCT)图像自动匹配的准确性。 CBCT图像是使用Catphan 504体模以各种总mAs比率(例如1.00、0.83、0.67、0.57和0.50)采集的。为了研究自动匹配精度,使用六维位移表获取了幻像图像。比较了图像质量和自动匹配校正。随着总mAs比率的降低,图像的噪声会增加,低对比度的分辨率会下降,而自动匹配的精度不会改变。因此,这项研究表明,获取CBCT图像时总mAs的变化对XI系统中Catphan 504体模的自动匹配没有影响.PACS编号:87.57.C-,87.57.Q-

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号