首页> 美国卫生研究院文献>Journal of Applied Clinical Medical Physics >Radiation dose and image quality comparison during spine surgery with two different intraoperative 3D imaging navigation systems
【2h】

Radiation dose and image quality comparison during spine surgery with two different intraoperative 3D imaging navigation systems

机译:使用两种不同的术中3D成像导航系统进行脊柱手术期间的辐射剂量和图像质量比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Careful protocol selection is required during intraoperative three‐dimensional (3D) imaging for spine surgery to manage patient radiation dose and achieve clinical image quality. Radiation dose and image quality of a Medtronic O‐arm commonly used during spine surgery, and a Philips hybrid operating room equipped with XperCT C‐arm 3D cone‐beam CT (hCBCT) are compared. The mobile O‐arm (mCBCT) offers three different radiation dose settings (low, standard, and high), for four different patient sizes (small, medium, large, and extra large). The patient's radiation dose rate is constant during the entire 3D scan. In contrast, C‐CBCT spine imaging uses three different field of views (27, 37, and 48 cm) using automatic exposure control (AEC) that modulates the patient's radiation dose rate during the 3D scan based on changing patient thickness. hCBCT uses additional x‐ray beam filtration. Small, medium, and large trunk phantoms designed to mimic spine and soft tissue were imaged to assess radiation dose and image quality of the two systems. The estimated measured “patient” dose for the small, medium, and large phantoms imaged by the mCBCT considering all the dose settings ranged from 9.4–27.6 mGy, 8.9–33.3 mGy, and 13.8–40.6 mGy, respectively. The “patient” dose values for the same phantoms imaged with hCBCT were 2.8–4.6 mGy, 5.7–10.0 mGy, and 11.0–15.2 mGy. The CNR for the small, medium, and large phantoms was 2.9 to 3.7, 2.0 to 3.0, and 2.5 to 2.6 times higher with the hCBCT system, respectively. Hounsfield unit accuracy, noise, and uniformity of hCBCT exceeded the performance of the mCBCT; spatial resolution was comparable. Added x‐ray beam filtration and AEC capability achieved clinical image quality for intraoperative spine surgery at reduced radiation dose to the patient in comparison to a reference O‐arm system without these capabilities.
机译:在脊柱手术的术中三维(3D)成像期间,需要仔细选择方案,以管理患者的放射剂量并达到临床图像质量。比较了脊柱外科手术中常用的Medtronic O型臂和配备XperCT C型臂3D锥形束CT(hCBCT)的Philips混合手术室的辐射剂量和图像质量。移动式O型臂(mCBCT)提供三种不同的辐射剂量设置(低,标准和高),适用于四种不同的患者体型(小,中,大和特大)。在整个3D扫描过程中,患者的辐射剂量率是恒定的。相比之下,C-CBCT脊柱成像使用自动曝光控制(AEC)使用三个不同的视野(27、37和48厘米),该自动曝光控制(AEC)在3D扫描过程中根据患者的厚度变化来调制患者的辐射剂量率。 hCBCT使用额外的X射线束过滤。对用于模拟脊椎和软组织的小型,中型和大型躯干体模进行成像,以评估两个系统的辐射剂量和图像质量。考虑到所有剂量设置,通过mCBCT成像的小型,中型和大型体模的估计“患者”剂量估计值分别为9.4-27.6 mGy,8.9-33.3 mGy和13.8-40.6 mGy。用hCBCT成像的相同体模的“患者”剂量值为2.8–4.6 mGy,5.7–10.0 mGy和11.0–15.2 mGy。使用hCBCT系统,小,中和大体模的CNR分别高2.9到3.7、2.0到3.0和2.5到2.6倍。 hCBCT的Hounsfield单位准确性,噪声和均匀性超过了mCBCT的性能;空间分辨率是可比的。与没有这些功能的参考O型臂系统相比,增加的X射线束过滤和AEC功能可在减少对患者的辐射剂量的情况下实现术中脊柱外科手术的临床图像质量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号