Design of a digital beam attenuation system for computed tomography. Part II. Performance study and initial results

机译:用于计算机断层扫描的数字束衰减系统的设计。第二部分绩效研究和初步结果

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

>Purpose: The purpose of this work is to present a performance study of the digital beam attenuator (DBA) for implementing fluence field modulated CT (FFMCT) using a simulation framework developed to model the incorporation of the DBA into an existing CT system. Additionally, initial results will be presented using a prototype DBA and the realization of the prototype will be described. To our knowledge, this study represents the first experimental use of a device capable of modulating x-ray fluence as a function of fan angle using a CT geometry.>Methods: To realize FFMCT, the authors propose to use a wedge design in which one wedge is held stationary and another wedge is moved over the stationary wedge. Due to the wedge shape, the composite thickness of the two wedges changes as a function of the amount of overlap between the wedges. This design allows for the wedges to modulate the photon fluence incident onto a patient. Using a simulation environment, the effect of changing the number of wedges has on dose, scatter, detector dynamic range, and noise uniformity is explored. Experimental results are presented using a prototype DBA having ten Fe wedges and a c-arm CT system geometry. The experimental DBA results are compared to non-DBA scans using scatter and detector dynamic range as metrics. Both flat field and bowtie filtered CT acquisitions were simulated for comparison with the DBA.>Results: Numerical results suggest that substantial gains in noise uniformity and scatter-to-primary ratio (SPR) can be obtained using only seven wedges. After seven wedges, the decrease in noise ununiformity and SPR falls off at a lower rate. Simulations comparing CT acquisitions between flat field, bowtie enabled, and DBA CT acquisitions suggest DBA-FFMCT can reduce dose relative to flat field CT by ≈3 times. A bowtie filter under the same imaging conditions was shown to only allow a dose reduction of 1.65 times. Experimentally, a 10 wedge DBA prototype result showed a SPR reduction of ≈4 times relative to flat field CT. The dynamic range for the DBA prototype was 3.7 compared to 84.2 for the flat field scan.>Conclusions: Based on the results presented in this paper and the companion paper [T. Szczykutowicz and C. Mistretta, “Design of a digital beam attenuation system for computed tomography. Part I. System design and simulation framework,” Med. Phys. 40, 021905 (2013)] doi: 10.1118/1.4773879., FFMCT implemented via the DBA device seems feasible and should result in both a dose reduction and an improvement in image quality as judged by noise uniformity and scatter reduction. In addition, the dynamic range reduction achievable using the DBA may allow photon counting imaging to become a clinical reality. This study may allow for yet another step to be taken in the field of patient specific dose modulation.
机译:>目的:这项工作的目的是介绍一种数字束衰减器(DBA)的性能研究,该数字束衰减器使用开发用于模拟将DBA并入其中的仿真框架来实现通量场调制CT(FFMCT)。现有的CT系统。此外,将使用原型DBA展示初始结果,并描述原型的实现。据我们所知,这项研究代表了一种首次实验性使用的装置,该装置能够使用CT几何形状根据扇形角度调节X射线通量。>方法:为了实现FFMCT,作者建议使用楔形设计,其中一个楔形保持固定,另一个楔形在固定楔形上移动。由于楔形,两个楔的复合厚度根据楔之间的重叠量而变化。这种设计允许楔形物调节入射到患者身上的光子通量。在仿真环境中,探讨了改变楔形数量对剂量,散射,检测器动态范围和噪声均匀性的影响。使用具有十个Fe楔形和c型臂CT系统几何形状的原型DBA展示了实验结果。使用散射和检测器动态范围作为度量,将实验DBA结果与非DBA扫描进行比较。模拟了平场和领结滤波后的CT采集,以与DBA进行比较。>结果:数值结果表明,仅使用七个就可以获得噪声均匀性和散射本原率(SPR)的实质性提高。楔子。经过七个楔形后,噪声不均匀性和SPR的降低以较低的速率下降。模拟比较了平场,领结使能和DBA CT采集之间的CT采集,表明DBA-FFMCT可以将相对于平场CT的剂量降低约3倍。在相同成像条件下的领结式过滤器仅能减少1.65倍的剂量。实验上,一个10楔形DBA原型结果显示,相对于平场CT,SPR降低了约4倍。 DBA原型的动态范围为3.7,而平场扫描的动态范围为84.2。>结论:基于本文和随附论文[T. C. S.]的结果。 Szczykutowicz和C. Mistretta,“用于计算机断层扫描的数字束衰减系统的设计。第一部分:系统设计和仿真框架。物理40,021905(2013)],doi:10.1118 / 1.4773879。,通过DBA设备实施的FFMCT似乎是可行的,并且应降低剂量,并通过噪声均匀性和散射降低来判断图像质量的提高。另外,使用DBA可实现的动态范围减小可允许光子计数成像成为临床现实。这项研究可能允许在患者特定剂量调节领域采取另一步骤。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号