Biomechanical analysis of cervical range of motion and facet contact force after a novel artificial cervical disc replacement

机译:新型人工颈椎间盘置换术后颈椎活动范围和小平面接触力的生物力学分析

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

An ideal implantation of artificial cervical disc replacement (ACDR) prosthesis should preserve the cervical biomechanics and maintain normal cervical biomechanics. In this study, we designed a novel prosthesis based on the physiological curvature of the cervical endplate, and conducted an in vitro test with cadaveric cervical specimens to study its effect on cervical range of motion (ROM) and facet contact force. Eighteen cadaveric cervical specimens (C2-C7) were evaluated with a 50 N follower load and a moment of ± 2.0 N-m. Testing conditions were as follows: control (group 1), C5-C6 ACDR with a novel prosthesis (group 2), C5-C6 ACDR with the Prestige LP prosthesis (group 3) and C5-C6 cervical discectomy and fusion (ACDF) with cage internal fixation (group 4). The range of motion (ROM) of all segments and facet joint contact force were measured and analyzed. The results showed that there was no significant difference among the group 1, 2 and 3 in terms of ROM (P>0.05). The ROM of C5-C6 in the group 4 showed a significant decrease compared with the group 1, 2, and 3 (P<0.05). The group 2 had a similar facet joint force with the group 1, and there is no statistical difference among the group 1, 2 and 3 (P>0.05). The facet joint force of C5-C6 in the group 4 reduced significantly compared with the other groups (P<0.05). In conclusion, both novel cervical prosthesis and Prestige LP could better preserve the normal cervical ROM and maintain facet joint force than ACDF.
机译:理想的人工颈椎间盘置换术(ACDR)假体植入应保留子宫颈生物力学并维持正常的子宫颈生物力学。在这项研究中,我们基于宫颈终板的生理曲度设计了一种新型假体,并使用尸体宫颈标本进行了体外测试,以研究其对宫颈运动范围(ROM)和小平面接触力的影响。对18个尸体宫颈标本(C2-C7)进行了50 N的从动负载评估,其力矩为±2.0 N-m。测试条件如下:对照组(第1组),带有新型假体的C5-C6 ACDR(第2组),带有Prestige LP假体的C5-C6 ACDR(第3组),以及C5-C6颈椎间盘摘除融合术(ACDF)笼内固定(第4组)。测量并分析所有节段的运动范围(ROM)和小关节接触力。结果表明,第1、2和3组在ROM方面无显着差异(P> 0.05)。与第1、2和3组相比,第4组中C5-C6的ROM显示出显着降低(P <0.05)。第2组的小关节关节力与第1组相似,第1、2和3组之间无统计学差异(P> 0.05)。与其他组相比,第4组C5-C6的小关节力显着降低(P <0.05)。综上所述,新型颈椎假体和Prestige LP均比ACDF更好地保留了正常的颈ROM,并保持了小关节力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号