Achieving the Recommended Endotracheal Tube Cuff Pressure: A Randomized Control Study Comparing Loss of Resistance Syringe to Pilot Balloon Palpation

机译:实现推荐的气管插管袖带压:一项随机对照研究比较阻力注射器对飞行员触诊的损失

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Background Both under- and overinflation of endotracheal tube cuffs can result in significant harm to the patient. The optimal technique for establishing and maintaining safe cuff pressures (20–30 cmH2O) is the cuff pressure manometer, but this is not widely available, especially in resource-limited settings where its use is limited by cost of acquisition and maintenance. Therefore, anesthesia providers commonly rely on subjective methods to estimate safe endotracheal cuff pressure. This study set out to determine the efficacy of the loss of resistance syringe method at estimating endotracheal cuff pressures.
机译:背景技术气管插管袖带充气不足和过度充气都可能对患者造成重大伤害。建立和维持安全袖带压力(20–30 cmH2O)的最佳技术是袖带压力计,但这种方法尚不广泛,尤其是在资源有限的环境中,其使用受到购置和维护成本的限制。因此,麻醉提供者通常依靠主观方法来估计安全的气管内袖带压力。这项研究着手确定阻力损失注射器方法在估计气管内袖带压力方面的功效。