Urinary catheters are widely used for hospitalized patients and are often associated with high rates of urinary tract infection. We evaluated in vitro the antiadherence activity of a novel antiseptic Gendine-coated urinary catheter against several multidrug-resistant bacteria. Gendine-coated urinary catheters were compared to silver hydrogel-coated Foley catheters and uncoated catheters. Bacterial biofilm formation was assessed by quantitative culture and scanning electron microscopy. These data were further correlated to an in vivo rabbit model. We challenged 31 rabbits daily for 4 days by inoculating the urethral meatus with 1.0 × 109 CFU streptomycin-resistant Escherichia coli per day. In vitro, Gendine-coated urinary catheters reduced the CFU of all organisms tested for biofilm adherence compared with uncoated and silver hydrogel-coated catheters (P < 0.004). Scanning electron microscopy analysis showed that a thick biofilm overlaid the control catheter and the silver hydrogel-coated catheters but not the Gendine-coated urinary catheter. Similar results were found with the rabbit model. Bacteriuria was present in 60% of rabbits with uncoated catheters and 71% of those with silver hydrogel-coated catheters (P < 0.01) but not in those with Gendine-coated urinary catheters. No rabbits with Gendine-coated urinary catheters had invasive bladder infections. Histopathologic assessment revealed no differences in toxicity or staining. Gendine-coated urinary catheters were more efficacious in preventing catheter-associated colonization and urinary tract infections than were silver hydrogel-coated Foley catheters and uncoated catheters.
机译:导尿管被广泛用于住院患者,并且经常与尿路感染率高相关。我们在体外评估了一种新型杀菌剂甘地丁涂层导尿管对几种耐多药细菌的抗粘附活性。将涂有甘地丁的导尿管与涂有银水凝胶的Foley导尿管和未涂成的导尿管进行比较。通过定量培养和扫描电子显微镜评估细菌生物膜的形成。这些数据进一步与体内兔模型相关。我们每天向尿道口注入1.0×10 9 CFU链霉素抗性大肠杆菌,挑战31只兔子,持续4天。在体外,与未涂层和银水凝胶涂层导尿管相比,涂有甘地丁的导尿管降低了所有生物膜粘附性测试生物的CFU(P <0.004)。扫描电子显微镜分析表明,厚的生物膜覆盖在对照导管和涂有银水凝胶的导管上,但没有覆盖在甘丹碱上的导尿管。在兔子模型中发现了类似的结果。 60%的兔子在未涂导管的情况下存在细菌性尿病,在水银涂层的导管在兔子中的比例为71%(P <0.01),而在Gendine涂层的导尿管中则不存在。没有使用根丁胺涂层导尿管的兔子有浸润性膀胱感染。组织病理学评估显示毒性或染色无差异。与银水凝胶涂层的Foley导管和未涂层的导管相比,涂有金丁胺的导尿管在预防导管相关的定植和尿路感染方面更有效。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号