Can modern infrared analyzers replace gas chromatography to measure anesthetic vapor concentrations?

机译:现代红外分析仪能否代替气相色谱仪来测量麻醉剂蒸气浓度?

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

BackgroundGas chromatography (GC) has often been considered the most accurate method to measure the concentration of inhaled anesthetic vapors. However, infrared (IR) gas analysis has become the clinically preferred monitoring technique because it provides continuous data, is less expensive and more practical, and is readily available. We examined the accuracy of a modern IR analyzer (M-CAiOV compact gas IR analyzer (General Electric, Helsinki, Finland) by comparing its performance with GC.
机译:背景技术气相色谱法(GC)通常被认为是测量吸入麻醉蒸气浓度的最准确方法。但是,红外(IR)气体分析已成为临床上首选的监视技术,因为它可以提供连续的数据,更便宜,更实用并且易于使用。通过将其性能与GC进行比较,我们检查了现代红外分析仪(M-CAiOV紧凑型气体红外分析仪(通用电气,赫尔辛基,芬兰))的准确性。