首页> 美国卫生研究院文献>The British Journal of Ophthalmology >Use of a portable head mounted perimetry system to assess bedside visual fields
【2h】

Use of a portable head mounted perimetry system to assess bedside visual fields

机译:使用便携式头戴式视野检查系统评估床边视野

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

AIM—This study was designed to test the ability of a portable computer driven, head mounted visual field testing system to perform automated perimetry on patients at their bedside and to compare these results with the "gold standard" for bedside examinations, confrontation visual fields.
METHODS—The Kasha visual field system is a portable automated perimeter which utilises a virtual reality headset. 37 neurosurgery patients were examined at their bedside with a central 24 degree suprathreshold testing strategy after confrontation visual field testing. The patterns of visual field defects were categorised and compared with the results of confrontation testing.
RESULTS—A total of 42 field examinations were completed on 37 patients, and the average testing time for both eyes was 4.8 minutes with the perimetry system. Each of the 11 fields (100%) classified with defects on confrontation testing was similarly categorised on head mounted perimetry. 26 out of 31 (84%) visual fields were normal on both confrontation and perimetry testing, while five out of the 31 fields (16%) which were full on confrontation had visual field defects identified by head mounted perimetry.
CONCLUSION—The head mounted, automated perimetry system proved easily portable and convenient for examining neurosurgical patients at their bedside in the perioperative period. The device demonstrated equal sensitivity to confrontation visual field testing methods in detecting field defects and offers the advantage of standardised, quantifiable testing with graphic results for follow up examinations.

机译:目的:这项研究旨在测试便携式计算机驱动的头戴式视野测试系统对床旁患者进行自动视野检查的能力,并将这些结果与“金标准”进行床旁检查,对峙视野进行比较。
方法-Kasha视野系统是使用虚拟现实耳机的便携式自动外围设备。在对峙视野测试后,对37名神经外科患者在床旁进行了中心24度超阈值测试策略的检查。对视野缺损的模式进行分类,并与对抗测试的结果进行比较。
结果-对37位患者进行了总共42次现场检查,使用视野计系统对两只眼睛的平均检查时间为4.8分钟。在对抗测试中被分类为缺陷的11个场中的每一个(100%)都通过头戴式视野检查仪进行了类似的分类。在对抗和视野测试中,有31个视野中有26个(84%)正常,而在对抗中充满的31个视野中有5个(16%)具有通过头戴式视野检查法识别的视野缺陷。
结论—头戴式自动视野检查系统被证明易于携带,并且在围手术期在床旁检查神经外科患者非常方便。该设备在检测野外缺陷方面表现出与对抗性视野测试方法相同的敏感性,并具有标准化,可量化的测试优势以及图形化结果,可用于后续检查。