Comparison of matrix perimetry with octopus perimetry for assessing glaucomatous visual field defects.

机译:矩阵视野检查与章鱼视野检查的比较,以评估青光眼视野缺损。

获取原文
获取原文并翻译 | 示例
       

摘要

PURPOSE: To compare Humphrey Matrix (Matrix) with Octopus perimetry for assessing visual field (VF) indices, size, severity, and location in the glaucomatous VF defect. METHODS: Matrix (30-2 threshold program) and Octopus 1-2-3 perimetry (G1X program) were conducted on 60 eyes of 60 patients with primary open-angle glaucoma. The VF indices derived by Matrix (mean deviation, pattern standard deviation), and Octopus (mean defect, square root of loss variance, corrected loss variance) were compared, as were the size (percentage of points depressed < 5% or < 1%) and the severity (score and Hodapp-Parrish-Anderson grade) of the VF defect. RESULTS: Matrix and Octopus results agreed well overall for VF indices, size, severity, and locations of the VF defect. The mean deviation and pattern standard deviation in Matrix were significantly higher than the mean defect, square root of loss variance and square root of corrected loss variance in Octopus. On the total deviation plot, the size of the VF defect in the superior and nasal fields reported by Matrix was significantly higher than by Octopus. The VF defect score by Matrix was significantly higher than that by Octopus in the superior, nasal, and whole fields. More eyes had a worse grade by Matrix than by Octopus. CONCLUSIONS: Matrix compared well with Octopus in detecting glaucomatous visual field defects. However, the defects reported by Matrix were larger and more severe than with Octopus.
机译:目的:比较汉弗莱矩阵(矩阵)和章鱼视野,以评估视野(VF)指标,大小,严重程度和青光眼VF缺损的位置。方法:对60例原发性开角型青光眼患者的60只眼进行基质(30-2阈值程序)和章鱼1-2-3视野检查(G1X程序)。比较了由矩阵(均值偏差,模式标准偏差)和八达通(均值缺陷,损失方差的平方根,校正后的损失方差)得出的VF指数以及大小(压抑点的百分比<5%或<1%) )和VF缺陷的严重程度(分数和Hodapp-Parrish-Anderson评分)。结果:矩阵和章鱼的结果在VF指数,大小,严重性和VF缺陷的位置总体上一致。矩阵中的平均偏差和模式标准偏差显着高于章鱼的平均缺陷,损失方差的平方根和校正损失方差的平方根。在总偏差图上,Matrix报告的上,鼻腔中VF缺损的大小显着高于章鱼。在上,鼻和整个视野中,Matrix的VF缺陷评分均明显高于章鱼。 Matrix的评分比Octopus的评分差。结论:基质与章鱼相比在检测青光眼视野缺损方面具有良好的对比。但是,Matrix报告的缺陷比章鱼更大,更严重。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号