机译:Keeler Pulsair 2000非接触眼压计的重现性。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

AIMS--The IOP variation on repeated testing with the recently introduced Keeler Pulsair 2000 instrument was investigated. METHODS--One hundred normal individuals (50 male and 50 female) new to the instrument had three sets of IOP recordings within a 15 minute time period. RESULTS--The mean of the first set of IOPs from both right and left eyes was significantly higher than those from subsequent sets (p < 0.0001 for right eyes and p = 0.01 and < 0.0002 for left eyes). This tendency increased significantly with increasing IOP. Second and third IOP sets were, however, similar indicating stabilisation of IOP measurements. The coefficient of repeatability of the instrument between second and third sets was 4.2 mm Hg for right eyes and 3.6 mm Hg for left eyes. CONCLUSION--The Pulsair 2000 passes the British standard for reproducibility of a standard test method.
机译:目的-研究了使用最近推出的Keeler Pulsair 2000仪器进行重复测试后的IOP变化。方法-刚接触该仪器的100名正常个体(50例男性和50例女性)在15分钟的时间段内拥有三组IOP记录。结果-右眼和左眼的第一组IOP的平均值显着高于后续各组(右眼的p <0.0001,左眼的p = 0.01和<0.0002)。随着IOP的增加,这种趋势显着增加。但是,第二和第三套IOP相似,表明IOP测量值稳定。第二组和第三组仪器的重复性系数,右眼为4.2 mm Hg,左眼为3.6 mm Hg。结论-Pulsair 2000通过了英国标准测试方法的可重复性标准。