The applanation pneumatonograph combines pneumatic and electronic systems. It is used for measuring intraocular pressure, aqueous humour dynamics, and ocular pulsation. In the present study the PTG and the Goldmann tonometer have been used to measure the intraocular pressure in 40 normal and 40 glaucomatous eyes. It has been found that the results of the two instruments correlate well. The PTG readings have, generally, been found to be higher with a mean difference of 0-75 mmHg in the normal and 1-25 mmHg in the glaucomatous eye. The mean ocular pulse as determined in normal eyes was 3-25 +/-0-21 mmHg.
机译:压平式肺动脉造影仪结合了气动和电子系统。它用于测量眼内压,房水动力学和眼搏动。在本研究中,PTG和戈德曼眼压计已被用于测量40例正常和40例青光眼的眼内压。已经发现,两种仪器的结果具有良好的相关性。通常,PTG读数较高,正常眼和青光眼眼的平均差为0-75 mmHg,平均差为1-25 mmHg。在正常眼睛中确定的平均眼动脉冲为3-25 +/- 0-21 mmHg。