The rotating hinge knee prosthesis is a popular intervention in patients lacking stability with highly constrained total knee arthroplasty. Despite improvements in design, nonmechanical and mechanical complications continue to be a problem. Dislocation of the hinge has been widely described, mainly due to the component fracture. Few reports describe isolated dislocation of the rotating stem. We report a case of isolated disengagement of the rotating hinge mechanism, due to severe flexion gap imbalance, leading to subsequent posterior dislocation of the hinge and anterior knee dislocation, in a patient with a history of multiple total knee arthroplasty revisions. This case suggests the importance of the soft tissue balancing, the adequate patellar tracking, and use of a long cylindrical, minimally tapered rotating stem in hinge arthroplasty to minimize hinge dislocation.
机译:旋转铰链膝关节假体是对稳定性受到严重限制的全膝关节置换术患者的普遍干预。尽管设计有所改进,但非机械和机械复杂性仍然是一个问题。铰链的错位已被广泛描述,主要是由于组件断裂所致。很少有报道描述旋转茎的孤立脱位。我们报告了一个病例,该病例由于严重的屈曲间隙不平衡而导致旋转铰链机构孤立脱开,从而导致有多个全膝关节置换术修订史的患者发生铰链后脱位和前膝关节脱位。这种情况表明,在铰接关节置换术中平衡软组织,充分tracking骨跟踪以及使用长圆柱体,最小锥形旋转柄以最大程度地减少铰链脱位的重要性。