首页> 美国卫生研究院文献>Chronic Respiratory Disease >Cleaning and infection control of airway clearance devices used by CF patients
【2h】

Cleaning and infection control of airway clearance devices used by CF patients

机译:CF患者使用的气道清除装置的清洁和感染控制

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Respiratory treatment for cystic fibrosis (CF) patients includes use of respiratory devices. Contamination of airway clearance devices has not been adequately explored. We aimed to determine whether airway clearance devices are contaminated after use and whether cleaning guidelines for nebulizers are as effective for airway clearance devices. Patients brought their airway clearance devices to the clinic. Swabs from the devices were taken before and after cleaning and were cultured for bacterial counts. Total colony-forming units (CFU) was determined, and predominant colonies were identified using Matrix-Assisted Laser Desorption Ionization Time-of-Flight Mass Spectrometry technology. Thirty devices were collected from 23 patients. Most of the devices (28/30) were contaminated when brought to the clinic. Complete bacterial eradication was achieved in 15 (50%) samples and partial eradication in 9 (30%). The cleaning was totally ineffective in four samples. Median CFU decreased significantly from 1250 (IQR 25–75% 175–10.000) to 0 (IQR 25–75% 0–700) before and after cleaning (p < 0.0001). The predominant organisms were identified in five samples only, and there was no concordance with sputum culture results. Airway clearance devices are contaminated after use, and appropriate cleaning can reduce contamination. The effect on disease progression in CF patients is unclear. There is a need for infection prevention and control guidelines for the growing number of respiratory devices.
机译:囊性纤维化(CF)患者的呼吸治疗包括使用呼吸装置。尚未充分研究气道清理装置的污染。我们的目的是确定使用后气道清除装置是否受到污染,以及雾化器的清洁指南对于气道清除装置是否同样有效。患者将呼吸道清理设备带到诊所。在清洁之前和之后从设备中抽出棉签,并进行细菌计数培养。确定了总菌落形成单位(CFU),并使用基质辅助激光解吸电离飞行时间质谱技术鉴定了主要菌落。从23名患者中收集了30台设备。带到诊所时,大多数设备(28/30)被污染。 15个样本(50%)实现了彻底的细菌根除,而9个样本(30%)实现了部分根除。清洁对四个样品完全无效。清洁前后中位数CFU从1250(IQR 25–75%175–10.000)显着降低至0(IQR 25–75%0–700)(p <0.0001)。仅在五个样本中鉴定出主要微生物,并且与痰培养结果不一致。使用后,呼吸道清洁装置受到污染,适当的清洁可以减少污染。目前尚不清楚CF患者对疾病进展的影响。需要针对呼吸装置的数量不断增长的感染预防和控制指南。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号