Self-calibrating pulse contour cardiac output: do validation studies really show its clinical reliability?

机译:自校准脉冲轮廓心输出量:验证研究是否真的显示出其临床可靠性?

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The present study was performed to test a new software version of the FloTrac/Vigileo using head-up–head-down tilting in post-cardiac surgery patients. Impressive improvements in Bland and Altman limits of agreement from 37.5% to 21.6% were recorded. The results, however could be attributed to a failure to produce a wide enough range of test circulatory conditions. A more rigorous test of performance is needed before any real conclusion concerning use of the FloTrac/Vigileo in clinical practice can be made.
机译:进行本研究是为了对心脏手术后患者使用头朝上-头朝下倾斜的方式测试FloTrac / Vigileo的新软件版本。记录的布兰德和奥特曼协定范围从37.5%改善到21.6%,令人印象深刻。然而,结果可能归因于未能产生足够广泛的测试循环条件。在做出有关在临床实践中使用FloTrac / Vigileo的任何真实结论之前,需要对性能进行更严格的测试。