Effect of proximal contact strength on the three-dimensional displacements of implant-supported cantilever fixed partial dentures under axial loading

机译:近端接触强度对植入物支撑的悬臂固定局部义齿在轴向载荷下三维位移的影响

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Objective: This study investigated the effect of proximal contact strength on the three-dimensional displacements of cantilever fixed partial denture (CFPD) under vertically concentrated loading with digital laser speckle (DLS) technique. Methods: Fresh mandible of beagle dog was used to establish the implant-supported CFPD for specimen. DLS technique was employed for measuring the three-dimensional displacement of the prosthesis under vertically concentrated loading ranging from 200 to 3 000 g. The effect of the contact tightness on the displacement of CFPD was investigated by means of changing the contact tightness. Results: When an axial concentrated loading was exerted on the pontic of the implant-supported CFPD, the displacement of the CFPD was the greatest. The displacement of the prosthesis decreased with the increase of contact strength. When the contact strength was 0, 0.95, and 3.25 N, the displacement of the buccolingual direction was smaller than that of the mesiodistal direction but greater than that of the occlusogingival direction. When the force on the contact area was 6.50 N, the mesiodistal displacement of the prosthesis was the biggest while the buccolingual displacement was the smallest. Conclusions: The implant-supported CFPD is an effective therapy for fully or partially edentulous patients. The restoration of the contact area and the selection of the appropriate contact strength can reduce the displacement of the CFPD, and get a better stress distribution. The most appropriate force value is 3.25 N in this study.
机译:目的:本研究利用数字散斑技术(DLS)研究了垂直集中载荷下悬臂固定局部义齿(CFPD)三维位移对近端接触强度的影响。方法:用新鲜的比格犬下颌骨建立标本的植入物支持的CFPD。 DLS技术用于在200至3000 g的垂直集中载荷下测量假体的三维位移。通过改变接触紧密度,研究了接触紧密度对CFPD位移的影响。结果:当轴向集中载荷施加在植入物支撑的CFPD的桥体上时,CFPD的位移最大。假体的位移随着接触强度的增加而减小。当接触强度为0、0.95和3.25 N时,颊舌方向的位移小于近中颌骨方向的位移,但大于咬合龈方向的位移。当作用在接触区域的力为6.50 N时,假体的近中位移位最大,颊舌移位最小。结论:植入物支持的CFPD对完全或部分缺牙的患者是一种有效的治疗方法。接触面积的恢复和适当接触强度的选择可以减少CFPD的位移,并获得更好的应力分布。在这项研究中,最合适的力值为3.25N。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号