Projected Improvements in Accelerated Partial Breast Irradiation Using a Novel Breast Stereotactic Radiotherapy Device: A Dosimetric Analysis

机译:使用新型乳房立体定向放射治疗设备对加速局部乳房照射的预计改善:剂量分析

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Accelerated partial breast irradiation has caused higher than expected rates of poor cosmesis. At our institution, a novel breast stereotactic radiotherapy device has demonstrated dosimetric distributions similar to those in brachytherapy. This study analyzed comparative dose distributions achieved with the device and intensity-modulated radiation therapy accelerated partial breast irradiation. Nine patients underwent computed tomography simulation in the prone position using device-specific immobilization on an institutional review board–approved protocol. Accelerated partial breast irradiation target volumes (planning target volume_10mm) were created per the National Surgical Adjuvant Breast and Bowel Project B-39 protocol. Additional breast stereotactic radiotherapy volumes using smaller margins (planning target volume_3mm) were created based on improved immobilization. Intensity-modulated radiation therapy and breast stereotactic radiotherapy accelerated partial breast irradiation plans were separately generated for appropriate volumes. Plans were evaluated based on established dosimetric surrogates of poor cosmetic outcomes. Wilcoxon rank sum tests were utilized to contrast volumes of critical structures receiving a percentage of total dose (Vx). The breast stereotactic radiotherapy device consistently reduced dose to all normal structures with equivalent target coverage. The ipsilateral breast V20-100 was significantly reduced (P < .05) using planning target volume_10mm, with substantial further reductions when targeting planning target volume_3mm. Doses to the chest wall, ipsilateral lung, and breast skin were also significantly lessened. The breast stereotactic radiotherapy device’s uniform dosimetric improvements over intensity-modulated accelerated partial breast irradiation in this series indicate a potential to improve outcomes. Clinical trials investigating this benefit have begun accrual.
机译:局部乳房照射加速导致美容不良率高于预期。在我们的机构中​​,一种新颖的乳房立体定向放射治疗仪已经证明了与近距离放射治疗相似的剂量分布。这项研究分析了使用该设备获得的比较剂量分布,并且强度调制的放射疗法加速了部分乳房的照射。九名患者使用特定于设备的固定装置,通过机构审查委员会批准的方案,在俯卧位进行了计算机断层扫描模拟。根据国家外科手术辅助乳房和肠项目B-39协议,创建了加速的局部乳房照射目标体积(计划目标体积为10mm)。基于改善的固定性,使用较小的余量(计划目标体积为3mm)增加了乳房立体定向放射治疗的体积。对于适当的体积,分别生成了调强放射疗法和乳房立体定向放射疗法加速的局部乳房照射计划。根据已确立的不良化妆品效果的剂量替代指标评估计划。 Wilcoxon秩和检验用于对比接受一定百分比总剂量(Vx)的关键结构的体积。乳房立体定向放射治疗设备可始终如一地减少对所有正常结构的剂量,并具有相等的靶标覆盖率。使用计划目标体积为10mm时,同侧乳房V20-100明显降低了(P <.05),而当计划目标体积为3mm时,则有了进一步的降低。胸壁,同侧肺和乳房皮肤的剂量也明显减少。与本系列中的强度调制加速局部乳房照射相比,乳房立体定向放射疗法设备的剂量学一致性得到了改善,这表明可能会改善治疗效果。研究这种益处的临床试验已经开始产生。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号