A novel R-peak detection algorithm suitable for wearable electrocardiogram (ECG) devices is proposed with four objectives: robustness to noise, low latency processing, low resource complexity, and automatic tuning of parameters. The approach is a two-pronged algorithm comprising (1) triangle template matching to accentuate the slope information of the R-peaks and (2) a single moving average filter to define a dynamic threshold for peak detection. The proposed algorithm was validated on eight ECG public databases. The obtained results not only presented good accuracy, but also low resource complexity, all of which show great potential for detection R-peaks in ECG signals collected from wearable devices.
机译:提出了一种适用于可穿戴心电图(ECG)设备的新颖R峰检测算法,该算法具有四个目标:对噪声的鲁棒性,低延迟处理,低资源复杂性以及参数的自动调整。该方法是两管齐下的算法,包括(1)三角形模板匹配以强调R峰的斜率信息,以及(2)单个移动平均滤波器以定义用于峰值检测的动态阈值。该算法在8个ECG公共数据库上得到了验证。所获得的结果不仅具有良好的准确性,而且具有较低的资源复杂性,所有这些都显示了检测可穿戴设备收集的ECG信号中R峰的巨大潜力。