A Cuffless Blood Pressure Measurement Based on the Impedance Plethysmography Technique

机译:基于阻抗容积描记法的无袖血压测量

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

In the last decade, cuffless blood pressure measurement technology has been widely studied because it could be applied to a wearable apparatus. Electrocardiography (ECG), photo-plethysmography (PPG), and phonocardiography are always used to detect the pulse transit time (PTT) because the changed tendencies of the PTT and blood pressure have a negative relationship. In this study, the PPG signal was replaced by the impedance plethysmography (IPG) signal and was used to detect the PTT. The placement and direction of the electrode array for the IPG measurement were discussed. Then, we designed an IPG ring that could measure an accurate IPG signal. Twenty healthy subjects participated in this study. The changes in blood pressure after exercise were evaluated through the changes of the PTT. The results showed that the change of the systolic pressure had a better relationship with the change of the PTTIPG than that of the PTTPPG (r = 0.700 vs. r = 0.450). Moreover, the IPG ring with spot electrodes would be more suitable to develop with the wearable cuffless blood pressure monitor than the PPG sensor.
机译:在过去的十年中,无袖血压测量技术已经得到了广泛的研究,因为它可以应用于可穿戴设备。心电图(ECG),光电容积描记法(PPG)和心动描记法始终用于检测脉搏传播时间(PTT),因为PTT的变化趋势与血压呈负相关。在这项研究中,PPG信号被阻抗体积描记法(IPG)信号代替,并用于检测PTT。讨论了用于IPG测量的电极阵列的位置和方向。然后,我们设计了可以测量准确IPG信号的IPG环。二十名健康受试者参加了这项研究。运动后的血压变化通过PTT的变化进行评估。结果表明,收缩压的变化与PTTIPG的变化具有比PTTPPG更好的关系(r = 0.700 vs. r = 0.450)。此外,与PPG传感器相比,带有点电极的IPG环更适合与可穿戴式无袖血压计一起开发。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号