In this study, prototype ultra-thin fiber-optic dosimeters were fabricated using organic scintillators, wavelength shifting fibers, and plastic optical fibers. The sensor probes of the ultra-thin fiber-optic dosimeters consisted of very thin organic scintillators with thicknesses of 100, 150 and 200 μm. These types of sensors cannot only be used to measure skin or surface doses but also provide depth dose measurements with high spatial resolution. With the ultra-thin fiber-optic dosimeters, surface doses for gamma rays generated from a Co-60 therapy machine were measured. Additionally, percentage depth doses in the build-up regions were obtained by using the ultra-thin fiber-optic dosimeters, and the results were compared with those of external beam therapy films and a conventional fiber-optic dosimeter.
机译:在这项研究中,原型超薄光纤剂量计是使用有机闪烁体,波长偏移光纤和塑料光纤制造的。超薄光纤剂量计的传感器探头由厚度为100、150和200μm的非常薄的有机闪烁体组成。这些类型的传感器不仅可以用于测量皮肤或表面剂量,还可以提供具有高空间分辨率的深度剂量测量。使用超薄光纤剂量计,测量了从Co-60治疗机产生的伽马射线的表面剂量。另外,通过使用超薄光纤剂量计获得堆积区域中的深度剂量百分比,并将该结果与外部束治疗膜和常规光纤剂量计的结果进行比较。