Comparison of a Bioelectrical Impedance Device against the Reference Method Dual Energy X-Ray Absorptiometry and Anthropometry for the Evaluation of Body Composition in Adults

机译:生物电阻抗设备与参考方法的比较双能量X射线吸收和人体测量法评估成人的身体成分

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

This study aimed to compare the use of the bioelectrical impedance device (BIA) seca® mBCA 515 using dual X-ray absorptiometry (DXA) as a reference method, for body composition assessment in adults across the spectrum of body mass indices. It explores the utility of simple anthropometric measures (the waist height ratio (WHtR) and waist circumference (WC)) for the assessment of obesity. In the morning after an overnight fast (10 h), 30 participants underwent a body composition DXA (GE iDXA) scan, BIA (seca 515), and anthropometric measures. Compared to the DXA reference measure, the BIA underestimated fat mass (FM) by 0.32 kg (limits of agreement −3.8 kg, 4.4 kg); overestimated fat free mass (FFM) by 0.43 kg (limits of agreement −8.2 kg, 4.3 kg). Some of the variation was explained by body mass index (BMI), as for FM, the mean difference of the normal range BMI group was smaller than for the overweight/obese group (0.25 kg and 0.35 kg, respectively) with wider limits of agreement (−4.30 kg, 4.81 kg, and −3.61 kg, 4.30 kg, respectively). There were significant differences in visceral adipose tissue (VAT) volume measurements between methods with BIA systematically overestimating VAT compared to DXA. WC was more strongly correlated with DXA FM (rho = 0.90, p < 0.001) than WHtR (rho = 0.83, p < 0.001). BIA had some agreement with DXA; however, they are not equivalent measures for the range of BMIs explored, with DXA remaining the more informative tool. WC is a useful and simple assessment tool for obesity.
机译:这项研究旨在比较使用生物电阻抗设备(BIA)seca ® mBCA 515(使用双X射线吸收法(DXA)作为参考方法)对成人的身体成分评估体重指数。它探讨了简单的人体测量学方法(腰高比(WHtR)和腰围(WC))在评估肥胖症中的作用。在通宵禁食(10小时)后的早晨,有30位参与者进行了DXA(GE iDXA)身体扫描,BIA(seca 515)和人体测量。与DXA参考指标相比,BIA低估了脂肪量(FM)0.32千克(协议极限-3.8千克,4.4千克);高估了0.43公斤的无脂脂肪(FFM)(协议限制-8.2公斤,4.3公斤)。某些差异可以通过体重指数(BMI)进行解释,就FM而言,正常范围BMI组的平均差异小于超重/肥胖组(分别为0.25 kg和0.35 kg),并且一致范围更广(分别为-4.30公斤,4.81公斤和-3.61公斤,4.30公斤)。与DXA相比,BIA系统高估VAT的方法之间的内脏脂肪组织(VAT)体积测量值存在显着差异。 WC与DXA FM(rho = 0.90,p <0.001)的相关性比WHtR(rho = 0.83,p <0.001)更强。 BIA与DXA达成了一些协议;但是,对于所探究的BMI而言,它们并不是等效的衡量标准,而DXA仍然是提供更多信息的工具。 WC是肥胖的有用且简单的评估工具。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号