首页> 外文OA文献 >Comparison of a Bioelectrical Impedance Device against the Reference Method Dual Energy X-Ray Absorptiometry and Anthropometry for the Evaluation of Body Composition in Adults
【2h】

Comparison of a Bioelectrical Impedance Device against the Reference Method Dual Energy X-Ray Absorptiometry and Anthropometry for the Evaluation of Body Composition in Adults

机译:生物电阻抗装置对参考方法双能X射线吸收测定法和人体测量法的比较成人身体组成的评价

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

This study aimed to compare the use of the bioelectrical impedance device (BIA) seca® mBCA 515 using dual X-ray absorptiometry (DXA) as a reference method, for body composition assessment in adults across the spectrum of body mass indices. It explores the utility of simple anthropometric measures (the waist height ratio (WHtR) and waist circumference (WC)) for the assessment of obesity. In the morning after an overnight fast (10 h), 30 participants underwent a body composition DXA (GE iDXA) scan, BIA (seca 515), and anthropometric measures. Compared to the DXA reference measure, the BIA underestimated fat mass (FM) by 0.32 kg (limits of agreement −3.8 kg, 4.4 kg); overestimated fat free mass (FFM) by 0.43 kg (limits of agreement −8.2 kg, 4.3 kg). Some of the variation was explained by body mass index (BMI), as for FM, the mean difference of the normal range BMI group was smaller than for the overweight/obese group (0.25 kg and 0.35 kg, respectively) with wider limits of agreement (−4.30 kg, 4.81 kg, and −3.61 kg, 4.30 kg, respectively). There were significant differences in visceral adipose tissue (VAT) volume measurements between methods with BIA systematically overestimating VAT compared to DXA. WC was more strongly correlated with DXA FM (rho = 0.90, p < 0.001) than WHtR (rho = 0.83, p < 0.001). BIA had some agreement with DXA; however, they are not equivalent measures for the range of BMIs explored, with DXA remaining the more informative tool. WC is a useful and simple assessment tool for obesity.
机译:本研究旨在将生物电阻抗装置(BIA)SECA®MBCA515使用双X射线吸收计(DXA)作为参考方法进行比较,用于体重指数的成人体内的身体成分评估。它探讨了用于评估肥胖症的简单人类测量措施(腰部高比(WHTR)和腰围(WC))的效用。在一夜之间快速(10小时)后,30名参与者接受了身体成分DXA(GE IDXA)扫描,BIA(SECA 515)和人体测量措施。与DXA参考度量相比,BIA低估脂肪质量(FM)0.32千克(协议限值-3.8千克,4.4千克);高估的脂肪量(FFM)×0.43千克(协议限值-8.2千克,4.3千克)。通过体重指数(BMI)解释了一些变异,至于FM,正常范围BMI组的平均差异小于超重/肥胖基团(分别为0.25千克和0.35千克),具有更广泛的协议限制(-4.30千克,4.81千克,分别为-3.61千克,4.30千克)。与DXA相比,BIA系统高估增值税的方法之间的内脏脂肪组织(VAT)体积测量存在显着差异。 WC与DXA FM(RHO = 0.90,P <0.001)更强烈地相关(RHO = 0.83,P <0.001)。比亚与DXA一致;但是,它们对探索BMI的范围不等同的措施,DXA剩下更具信息性的工具。 WC是肥胖的有用而简单的评估工具。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号