Validation of a Liquid Chromatography-Electrospray Ionization-Tandem Mass Spectrometry Method for Determination of All-Trans Retinoic Acid in Human Plasma and Its Application to a Bioequivalence Study

机译:液相色谱-电喷雾串联质谱法测定人血浆中全反式维甲酸的方法的验证及其在生物等效性研究中的应用

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A sensitive, reliable and specific LC-MS-MS method was developed and validated for the identification and quantitation of all-trans retinoic acid (ATRA) in human plasma. Acitretin was used as the internal standard (IS). After liquid-liquid extraction of 500 μL plasma with methyl tert-butyl ether (MTBE), ATRA and the IS were chromatographed on a HyPURITY C18 column (150 mm × 2.1 mm, 5 μm) with the column temperature set at 40 °C. The mobile phase was consisted of 40% phase A (MTBE–methanol–acetic acid, 50:50:0.5, v/v) and 60% phase B (water–methanol–acetic acid, 50:50:0.5, v/v) with a flow rate of 0.3 mL/min. The API 4000 triple quadrupole mass spectrometer was operated in multiple reaction monitoring (MRM) mode via the positive electrospray ionization interface using the transition m/z 301.4 → 123.1 for ATRA and m/z 326.9 → 177.1 for IS, respectively. The calibration curve was linear over the range of 0.45–217.00 ng/mL (r ≥ 0.999) with a lower limit of quantitation (LLOQ) of 0.45 ng/mL. The intra- and inter-day precisions values were below 8% relative standard deviation and the accuracy was from 98.98% to 106.19% in terms of relative error. The validated method was successfully applied in a bioequivalence study of ATRA in Chinese healthy volunteers.
机译:开发了一种灵敏,可靠且特异的LC-MS-MS方法,并已用于鉴定和定量人血浆中的全反式维甲酸(ATRA)。阿维A酸用作内标(IS)。用甲基叔丁基醚(MTBE)液-液萃取500μL血浆后,在HyPURITY C18色谱柱(150 mm×2.1 mm,5μm)上色谱分离ATRA和IS,色谱柱温度设定为40°C。流动相由40%A相(MTBE-甲醇-乙酸,50:50:0.5,v / v)和60%B相(水-甲醇-乙酸,50:50:0.5,v / v)组成),流速为0.3 mL / min。 API 4000三重四极杆质谱仪通过正电喷雾电离界面在多反应监测(MRM)模式下运行,分别对ATRA使用m / z 301.4→123.1,对于IS使用m / z 326.9→177.1。校正曲线在0.45–217.00 ng / mL范围内呈线性(r≥0.999),定量下限(LLOQ)为0.45 ng / mL。日内和日间精度值相对标准偏差低于8%,相对误差的准确度从98.98%到106.19%。验证的方法已成功应用于中国健康志愿者的ATRA生物等效性研究。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号