Fracture of a Polyethylene Tibial Post in a Scorpio Posterior-Stabilized Knee Prosthesis

机译:天蝎后稳定型膝关节假体中聚乙烯胫骨桩的骨折

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

We report the case of a polyethylene tibial post fracture in a 72-year-old woman 14 months after a Scorpio posterior-stabilized (PS) total knee arthroplasty. The polyethylene wear was found around the fracture site of the post, especially over the anterior aspect of the post base. The failure mechanism of the post fracture in the present case was anterior impingement with excessive wear over the base of the anterior aspect of the tibial post, which became a stress-riser of post and cam articulation. This is the first report of a polyethylene tibial post fracture of a Scorpio PS prosthesis.
机译:我们报道了一名72岁的女性在胫骨后路稳定(PS)全膝关节置换术后14个月发生聚乙烯胫骨骨折的情况。在支柱的断裂部位周围发现了聚乙烯磨损,特别是在支柱底部的前部。在本例中,骨折后的破坏机制是前部撞击,胫骨桩前部的基部过度磨损,这成为桩和凸轮关节应力的增加。这是Scorpio PS假体的聚乙烯胫骨骨折后的首次报道。