首页> 美国卫生研究院文献>The Malaysian Journal of Medical Sciences : MJMS >Determination of Tube Output (kVp) and Exposure Mode for Breast Phantom of Various Thicknesses/Glandularity for Digital Mammography
【2h】

Determination of Tube Output (kVp) and Exposure Mode for Breast Phantom of Various Thicknesses/Glandularity for Digital Mammography

机译:乳腺X线摄影术不同厚度/腺体的乳腺体模的管输出(kVp)和曝光模式的确定

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Background: Optimisation of average glandular dose (AGD) for two-dimensional (2D) mammography is important, as imaging using ionizing radiation has the probability to induce cancer resulting from stochastic effects. This study aims to observe the effects of kVp, anode/filter material, and exposure mode on the dose and image quality of 2D mammography.Methods: This experimental study was conducted using full-field digital mammography. The entrance surface air kerma was determined using thermoluminescent dosimeter (TLD) 100H and ionization chamber (IC) on three types of Computerized Imaging Reference System (CIRS) phantom with 50/50, 30/70, and 20/80 breast glandularity, respectively, in the auto-time mode and auto-filter mode. The Euref protocol was used to calculate the AGD while the image quality was evaluated using contrast-to-noise ratio (CNR), figure of merit (FOM), and image quality figure (IQF).Results: It is shown that AGD values in the auto-time mode did not decrease significantly with the increasing tube voltage of the silver filter (r = −0.187, P > 0.05) and rhodium filter (r = −0.131, P > 0.05) for all the phantoms. The general linear model showed that AGD for all phantoms had a significant effect between different exposure factors [F (6,12.3) = 4.48 and mode of exposure F (1,86) = 4.17, P < 0.05, respectively] but there is no significant difference between the different anode/filter combination [F (1,4) = 0.571].Conclusion: In summary, the 28, 29, and 31 kVp are the optimum kVp for 50%, 30%, and 20% breast glandularity, respectively. Besides the auto-filter mode is suitable for 50%, 30%, and 20% breast glandularity because it is automatic, faster, and may avoid error done by the operator.
机译:背景:二维(2D)乳房X光检查的平均腺体剂量(AGD)的优化非常重要,因为使用电离辐射成像具有诱发随机效应导致癌症的可能性。本研究旨在观察kVp,阳极/滤光片材料和曝光方式对2D乳腺摄影的剂量和图像质量的影响。方法:本实验研究是使用全场数字乳腺摄影进行的。使用热剂量计(TLD)100H和电离室(IC)在分别具有50 / 50、30 / 70和20/80乳腺腺体的三种类型的计算机成像参考系统(CIRS)体模上确定入射表面空气比释动能,在自动时间模式和自动过滤模式下。使用Euref协议计算AGD,同时使用对比噪声比(CNR),品质因数(FOM)和图像质量图(IQF)评估图像质量。结果:显示AGD值在对于所有体模,自动时间模式均不会随着银滤镜(r = -0.187,P> 0.05)和铑滤镜(r = -0.131,P> 0.05)的管电压的增加而显着降低。总体线性模型表明,所有体模的AGD在不同的暴露因子[F(6,12.3)= 4.48和暴露模式F(1,86)= 4.17,P <0.05)之间均具有显着影响,但没有结论:总而言之,对于50%,30%和20%的乳腺腺体,28、29和31 kVp是最佳kVp,总而言之,阳极/过滤器组合之间的显着差异[F(1,4)= 0.571]。分别。此外,自动过滤模式还适合于50%,30%和20%的乳腺腺体,因为它是自动的,更快的,并且可以避免操作员犯错误。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号