Mini C-Arm Fluoroscopy: Does Its Configuration Matter for Radiation Exposure to the Surgeon?

机译:小型C型臂荧光透视术:其配置对放射给外科医生而言是否重要?

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

>Background: The risk of occupational radiation exposure to the surgeon associated with the use of a mini C-arm has yet to reach a wide consensus. Using a distal radius fracture surgery model, we tested the hypothesis that radiation exposure to the surgeon’s critical body parts is independent of mini C-arm configuration. >Methods: An anthropomorphic mannequin (representing the upper body of a 60” male surgeon) was seated at a hand table as if operating on a volar-plated wrist Sawbone model. Thermoluminescent dosimeters measured radiation exposure to the surgeon’s eyes, thyroid, chest, hand, and groin from a mini C-arm fluoroscopy unit in 3 commonly used configurations: vertical (source above table), inverted (source below table), and horizontal (with beam parallel to table surface). The fluoroscope scanned the wrist model for 15 continuous minutes in triplicate for each orientation. >Results: Radiation to the hand was significantly greatest in all mini C-arm positions compared with all other anatomic sites irrespective of C-arm position. Hand radiation exposure was greatest in the horizontal position (2887.09 mrem), versus the vertical and inverted positions (59.79 mrem, 31.10 mrem, P < .001). Eye radiation exposure was significantly greater in the inverted position (2.33 mrem) compared with the vertical (0.67 mrem, P = .024), and horizontal positions (0.33 mrem, P = .012). No significant difference in radiation exposure was found at the thyroid, chest, and groin sites, at each of the 3 C-arm configurations. >Conclusions: The model’s hand received almost 1000 times more radiation exposure than all other anatomic sites with statistically greatest radiation exposure sustained in the horizontal position. Eye radiation exposure with the C-arm in the inverted position (below the table) was also significantly greater.
机译:>背景:与使用迷你C型臂相关的外科医生职业性放射线暴露的风险尚未达成广泛共识。通过使用radius骨远端骨折手术模型,我们检验了以下假设:外科医生关键身体部位的辐射暴露与微型C臂配置无关。 >方法:一个拟人化的人体模型(代表一个60英寸的男性外科医生的上半身)坐在一个手掌上,就像在手掌上腕骨锯齿模型上操作一样。热发光剂量计在3种常用配置中测量了来自微型C型臂荧光透视仪对外科医生的眼睛,甲状腺,胸部,手和腹股沟的放射线暴露:垂直(上表中的光源),倒置(上表的光源)和水平(下光源)梁平行于工作台表面)。荧光镜对每个方向一式三份地连续15分钟扫描腕部模型。 >结果:与所有其他解剖部位相比,无论C臂的位置如何,在所有微型C臂位置上,手部的辐射均显着最大。与水平和垂直位置(59.79 mrem,31.10 mrem,P <.001)相比,水平位置(2887.09 mrem)的手辐射暴露最大。与垂直位置(0.67 mrem,P = .024)和水平位置(0.33 mrem,P = .012)相比,倒置位置(2.33 mrem)的眼睛辐射暴露明显更大。在3个C臂配置中,在甲状腺,胸部和腹股沟部位的辐射暴露均无显着差异。 >结论:该模型的手比其他解剖部位的辐射暴露要高出将近1000倍,在统计上,水平位置的辐射暴露最大。当C型臂处于倒置位置时(表下方),眼睛的辐射暴露也大大增加。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号