Effects of Hot Sterilization on Torsional Properties of Endodontic Instruments: Systematic Review with Meta-Analysis

机译:热灭菌对牙髓器械扭转特性的影响:荟萃分析的系统评价

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Sterilization is a fundamental step in the reuse of endodontic instruments. The sterilization procedure involves disinfection, cleaning, washing, drying, packaging, and sterilization by heat. Heat sterilization can lead to changes in the physical and mechanical properties of dental instruments. These changes can affect the external surfaces via micropitting, corrosion, a reduction in cutting capacity, and/or an influence on the resistance to cyclic fatigue or to torsional fatigue. In this study, we examined the modification of the torsional properties of endodontic instruments after hot sterilization, and compared the properties with instruments not subjected to hot sterilization cycles in terms of resistance to torsional fatigue and deflection angle in NiTi and steel instruments. The following work was performed based on the PRISMA indications. Studies were identified through bibliographic research using electronic databases. A total of 725 records were identified in the PubMed and Scopus databases. A total of 685 records remained after exclusion by year of publication (1979 to 2019). With the application of the eligibility criteria (all articles pertaining to the issue of sterilization in endodontics), we found 146 articles, which decreased to 130 articles after elimination of duplications. There were 45 articles that studied the influences of sterilization procedures on the physical and mechanical characteristics of the instruments, and 12 that measured parameters related to resistance to torsional fatigue. Applying the inclusion and exclusion criteria resulted in a total of eight articles for quantitative analysis. The meta-analysis results show a pejorative effect of torsional fatigue for NiTi instruments subjected to heat sterilization compared to the non-sterilized control.
机译:消毒是牙髓器械重复使用的基本步骤。消毒程序包括消毒,清洁,清洗,干燥,包装和加热消毒。加热灭菌会导致牙科器械的物理和机械性能发生变化。这些变化会通过微点蚀,腐蚀,切削能力降低和/或对循环疲劳或扭转疲劳的影响而影响外表面。在这项研究中,我们检查了热灭菌后牙髓器械的扭转特性的变化,并就镍钛合金和钢制器械的抗扭转疲劳性和挠曲角,将其与未经热灭菌循环的器械进行了比较。根据PRISMA适应症进行了以下工作。通过使用电子数据库的书目研究来鉴定研究。在PubMed和Scopus数据库中总共鉴定了725条记录。按出版年份(1979年至2019年)排除后,总共剩余685条记录。在符合资格标准的情况下(所有文章均涉及牙髓消毒的问题),我们发现146篇文章,在消除重复后减少到130篇。有45篇文章研究了灭菌程序对器械物理和机械特性的影响,还有12篇文章测量了与抗扭转疲劳性有关的参数。应用纳入和排除标准后,共有八篇文章进行了定量分析。荟萃分析结果显示,与未灭菌的对照相比,经过热灭菌的NiTi器械具有扭转疲劳的不良影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号