Image Quality of Digital Direct Flat-Panel Mammography Versus an Indirect Small-Field CCD Technique Using a High-Contrast Phantom

机译:数字直接平板乳腺X线摄影术与使用高对比度幻像的间接小视场CCD技术的图像质量

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Objective. To compare the detection of microcalcifications on mammograms of an anthropomorphic breast phantom acquired by a direct digital flat-panel detector mammography system (FPM) versus a stereotactic breast biopsy system utilizing CCD (charge-coupled device) technology with either a 1024 or 512 acquisition matrix (1024 CCD and 512 CCD). Materials and Methods. Randomly distributed silica beads (diameter 100–1400 μm) and anthropomorphic scatter bodies were applied to 48 transparent films. The test specimens were radiographed on a direct digital FPM and by the indirect 1024 CCD and 512 CCD techniques. Four radiologists rated the monitor-displayed images independently of each other in random order. Results. The rate of correct positive readings for the “number of detectable microcalcifications” for silica beads of 100–199 μm in diameter was 54.2%, 50.0% and 45.8% by FPM, 1024 CCD and 512 CCD, respectively. The inter-rater variability was most pronounced for silica beads of 100–199 μm in diameter. The greatest agreement with the gold standard was observed for beads >400 μm in diameter across all methods. Conclusion. Stereotactic spot images taken by 1024 matrix CCD technique are diagnostically equivalent to direct digital flat-panel mammograms for visualizing simulated microcalcifications >400 μm in diameter.
机译:目的。比较直接数字平板探测器X线乳腺摄影系统(FPM)与采用CCD(电荷耦合器件)技术和1024或512采集矩阵的立体定向乳腺活检系统采集的拟人化乳体模型乳房X线照片上微钙化的检测结果(1024 CCD和512 CCD)。材料和方法。将随机分布的二氧化硅珠(直径100-1400μm)和拟人散射体应用于48个透明膜。通过直接数字FPM以及间接1024 CCD和512 CCD技术对测试样品进行射线照相。四位放射线医师以随机顺序彼此独立地对监视器显示的图像进行评级。结果。对于FPM,1024 CCD和512 CCD,直径为100-199μm的硅珠“可检测到的微钙化次数”的正确阳性读数率分别为54.2%,50.0%和45.8%。粒径为100–199μm的硅珠的评估者间差异最为明显。在所有方法中,对于直径>400μm的珠子,均观察到与金标准的最大一致性。结论。通过1024矩阵CCD技术拍摄的立体定向点图像在诊断上等效于直接数字平板乳腺X光照片,以可视化直径>400μm的模拟微钙化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号