During the 2001 U. S. West Nile virus (WNV) season, 163 specimens were reactive in an in-house WNV-specific immunoglobulin M (IgM) screening enzyme-linked immunosorbent assay (ELISA) and were referred to either the Centers for Disease Control and Prevention or the appropriate state public health laboratory (CDC/SPHL) for additional testing. CDC/SPHL supplied results for 124 specimens that could be further evaluated in-house: 70 specimens were nonreactive in the CDC/SPHL WNV-specific IgM screening assay, and 54 specimens were reactive. These specimens were used to evaluate a modified in-house WNV-specific IgM ELISA that incorporated background subtraction to identify nonspecific reactivity and thus improve assay specificity. Of the 70 CDC/SPHL nonreactive samples, 49 (70%) were nonreactive in the modified ELISA; of the 54 CDC/SPHL reactive samples, 51 (94%) were reactive in the modified ELISA. Confirmatory studies performed by CDC/SPHL indicated that 38 CDC/SPHL screen-reactive specimens represented true WNV infection; all 38 specimens were reactive in the modified in-house WNV-specific IgM ELISA. These findings demonstrate that an in-house ELISA system for WNV-specific IgM effectively identifies patients with WNV infection.
机译:在2001年美国西尼罗河病毒(WNV)季节中,有163个标本在内部WNV特异性免疫球蛋白M(IgM)筛查酶联免疫吸附测定(ELISA)中具有反应性,并移交给了疾病控制和预防中心或适当的州公共卫生实验室(CDC / SPHL)进行其他测试。 CDC / SPHL提供了124个标本的结果,这些标本可以在内部进行进一步评估:70个标本在CDC / SPHL WNV特异性IgM筛选分析中不具有反应性,而54个标本具有反应性。这些标本用于评估经过修改的内部WNV特异性IgM ELISA,该方法结合了背景扣除法以鉴定非特异性反应性,从而提高测定特异性。在70个CDC / SPHL非反应性样品中,有49个(70%)在改良的ELISA中不反应。在54个CDC / SPHL反应性样品中,有51个(94%)在改良的ELISA中具有反应性。 CDC / SPHL进行的确证研究表明,有38个CDC / SPHL筛查反应性标本代表真正的WNV感染。在改良的内部WNV特异性IgM ELISA中,所有38个样本均具有反应性。这些发现表明,针对WNV特异性IgM的内部ELISA系统可有效识别WNV感染的患者。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号