首页> 外文期刊>Clinical and vaccine immunology: CVI >Performance of a Commercial Immunoglobulin M Antibody Capture Assay Using Analyte-Specific Reagents To Screen for Interfering Factors during a West Nile Virus Epidemic Season in Nebraska
【24h】

Performance of a Commercial Immunoglobulin M Antibody Capture Assay Using Analyte-Specific Reagents To Screen for Interfering Factors during a West Nile Virus Epidemic Season in Nebraska

机译:使用特定于分析物的试剂筛选内布拉斯加州西尼罗河病毒流行季节期间的干扰因素的商业免疫球蛋白M抗体捕获测定的性能

获取原文

摘要

In 2003, the Nebraska Public Health Laboratory tested more than 10,371 serum and 516 cerebral spinal fluid specimens. Results showed that without performing the interfering factors screen for specimens in the low positive index value range of >1.1 to ≤3.5, a false positivity rate of 6.5% would have been realized.
机译:2003年,内布拉斯加州公共卫生实验室检测了10,371份血清和516份脑脊髓液标本。结果表明,如果不对样本的低正指数值范围> 1.1至≤3.5进行干扰因素筛选,则可以实现6.5%的假阳性率。