Total 25-Hydroxyvitamin D Determination by an Entry Level Triple Quadrupole Instrument: Comparison between Two Commercial Kits

机译:入门级三重四极杆仪器测定的总25-羟维生素D:两种商业试剂盒之间的比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Objective. 25-hydroxyvitamin D2/D3 (25-OHD2/D3) determination is a reliable biomarker for vitamin D status. Liquid chromatography-tandem mass spectrometry was recently proposed as a reference method for vitamin D status evaluation. The aim of this work is to compare two commercial kits (Chromsystems and PerkinElmer) for 25-OHD2/D3 determination by our entry level LC-MS/MS. Design and Methods. Chromsystems kit adds an online trap column to an HPLC column and provides atmospheric pressure chemical ionization, isotopically labeled internal standard, and 4 calibrator points. PerkinElmer kit uses a solvent extraction and protein precipitation method. This kit can be used with or without derivatization with, respectively, electrospray and atmospheric pressure chemical ionization. For each analyte, there are isotopically labeled internal standards and 7 deuterated calibrator points. Results. Performance characteristics are acceptable for both methods. Mean bias between methods calculated on 70 samples was 1.9 ng/mL. Linear regression analysis gave an R 2 of 0.94. 25-OHD2 is detectable only with PerkinElmer kit in derivatized assay option. Conclusion. Both methods are suitable for routine. Chromsystems kit minimizes manual sample preparation, requiring only protein precipitation, but, with our system, 25-OHD2 is not detectable. PerkinElmer kit without derivatization does not guarantee acceptable performance with our LC-MS/MS system, as sample is not purified online. Derivatization provides sufficient sensitivity for 25-OHD2 detection.
机译:目的。 25-羟基维生素D2 / D3(25-OHD2 / D3)测定是维生素D状态的可靠生物标志物。最近提出了液相色谱-串联质谱法作为维生素D状态评估的参考方法。这项工作的目的是通过我们的入门级LC-MS / MS比较两种用于测定25-OHD2 / D3的商业试剂盒(Chromsystems和PerkinElmer)。设计和方法。 Chromsystems试剂盒在HPLC色谱柱上增加了一个在线捕集柱,并提供大气压化学电离,同位素标记的内标和4个校准点。 PerkinElmer试剂盒使用溶剂萃取和蛋白质沉淀法。该试剂盒可分别在电喷雾和大气压化学电离作用下进行衍生化或不衍生化。对于每种分析物,都有同位素标记的内标和7个氘代校准物点。结果。两种方法的性能特征都是可以接受的。在70个样品上计算得出的方法之间的平均偏差为1.9µng / mL。线性回归分析得出R 2 为0.94。只有在衍生化验选件中使用PerkinElmer试剂盒才能检测到25-OHD2。结论。两种方法都适合常规。 Chromsystems试剂盒最大程度地减少了手动制备样品的过程,仅需要沉淀蛋白质,但是在我们的系统中,无法检测到25-OHD2。由于没有在线纯化样品,没有衍生化的PerkinElmer试剂盒不能保证我们的LC-MS / MS系统具有可接受的性能。衍生化为25-OHD2检测提供了足够的灵敏度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号