Improving Ion Mobility Measurement Sensitivity byUtilizing Helium in an Ion Funnel Trap

机译:通过提高离子迁移率测量灵敏度在离子漏斗中利用氦气

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Ion mobility instruments that utilize nitrogen as buffer gas are often preceded by an ion trap and accumulation region that also uses nitrogen, and for different inert gases, no significant effects upon performance are expected for ion mobility spectrometry (IMS) of larger ions. However, we have observed significantly improved performance for an ion funnel trap upon adding helium; the signal intensities for higher m/z species were improved by more than an order of magnitude compared to using pure nitrogen. The effect of helium upon IMS resolving power was also studied by introducing a He/N2 gas mixture into the drift cell, and in some cases, a slight improvement was observed compared to pure N2. The improvement in signal can be largely attributed to faster and more efficient ion ejection into the drift tube from the ion funnel trap.
机译:利用氮气作为缓冲气体的离子迁移仪通常在离子阱和聚集区之前也使用氮气,对于不同的惰性气体,对于较大离子的离子迁移谱(IMS),预期不会对性能产生重大影响。但是,我们发现加入氦气后离子漏斗阱的性能大大提高;与使用纯氮相比,更高m / z物种的信号强度提高了一个数量级以上。还通过将He / N2气体混合物引入漂移池来研究氦气对IMS分辨能力的影响,在某些情况下,与纯N2相比,观察到轻微改善。信号的改善在很大程度上归因于离子从离子漏斗中更快,更有效地喷射到漂移管中。