首页> 中文期刊> 《海峡药学》 >高效液相色谱-质谱联用测定人血浆中奥美拉唑的浓度

高效液相色谱-质谱联用测定人血浆中奥美拉唑的浓度

         

摘要

目的:建立高效液相色谱-质谱联用法测定人血浆中奥美拉唑的浓度。方法以泮托拉唑为内标,血浆样品经乙腈沉淀、取上清稀释后,进样,经HPLC-MS/MS分析。采用Ultimate XB-C18(2.1 mm ×150 mm,3μm),流动相甲醇(0.1%甲酸)∶水(0.1%甲酸)=(70∶30,V/V);流速:0.2mL・ min -1,柱温:40℃,电喷雾离子源(ESI),以多离子反应监测方式(MRM)进行正离子监测,奥美拉唑和内标泮托拉唑的定量分析离子对分别为m/z 346.2→m/z 198.0和m/z 384.0→m/z 200.0。结果奥美拉唑的线性范围为5.00~2000.00μg・ L-1,定量下限为5.00μg・ L -1。方法回收率在96.4%~107.9%,日内和日间精密度(RSD)均小于15%。结论该方法操作简便、灵敏、准确、快速,适用于奥美拉唑的人体药代动力学研究。%ABSTRACT:OBJECTIVE To establish a high performance liquid chromatography tandem mass spectrometry ( HPLC-MS/MS) method for the determination of Omeprazole in human plasma .METHODS Human plasma sam-ples were precipitated with acetonitrile and determined by HPLC-MS/MS using pantoprazole as the internal stand-ard.The separation was carried out on a Ultimate XB-C18(2.1mm ×150mm,3μm) with a mobile phase of methanol-water (including 0.1%formic acid) (70∶30) and the flow rate was 0.2mL・ min-1.ESI was performed in the MRM positive mode with using target ions at m/z 346.2→m/z 198.0 ( Omeprazole ) and m/z 384.0→m/z 200.0 ( panto-prazole).RESULTS The calibration curve of Omeprazole was obtained in the range of 5.00~2000.00μg・ L-1 with good linearity ( r>0.99 ) .The LLOQ of Omeprazole was 5μg・ L-1 .The method recovery was between 96.4%and 107.9%,and the intra and inter day RSD were less than 15%.CONCLUSION This method is convenient , sensitive ,accurate and fast ,which is suitable for human pharmacokinetic studies of Omeprazole .

著录项

相似文献

  • 中文文献
  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号