首页> 中文期刊> 《放射学实践 》 >辐射剂量对头颅CT扫描图像质量的影响

辐射剂量对头颅CT扫描图像质量的影响

             

摘要

目的:探讨不同辐射剂量行颅脑CT扫描对影像质量的影响。方法:8例正常男性志愿者用110~140kV,180~360mAs分四组行轴位平扫;各影像按颗粒均匀性、脑灰白质细节、界面清晰度和有无射线硬化等积分评定质量。结果:共扫描192次,120kV、240mAs以上组影像质量均达优良;130kV、300mAs以上组优片率更高。结论:采用适当辐射剂量(>120kV、240mAs)行颅脑CT扫描能提高密度分辨力,增加SNR,改善影像质量。%Objective: To evaluate the effect of different radiation doses to imaging quality in the course of plain cranial CT scans.Methods:Eight normal male volunteers were divided into four groups,and each group was performed plain cranial axial CT scans with 110,120,130,140kV and 180 ~ 360mAs respectively. All images were scored at four degrees (excellent, good, poor, very poor) according to grain homogeneity, fine structure appearance of the grey and white matter, hardening of X rays and clearness of tissue interface. Results:The patients were totally scanned 192 times. The image quality with > 120kV and > 240mAs is excellent or good. The rate of superior image quality with ≥ 130kV and≥300mAs, moreover, was higher. Conclusion:Using radiation dose with > 120kV and > 240mAs, the density resolution and SNR may be increased and imaging quality be improved.

著录项

相似文献

  • 中文文献
  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号