研究了Bi^3+在碳糊电极上的阳极溶出伏安特性,结果表明:在1.0mol·L^-1的HCl溶液中,于-0.055V处出现一灵敏的溶出峰,电极反应过程是受吸附控制的电极过程。并考察了测定Bi^3+的最优化条件,在1.0mol·L^-1的HCl溶液中,于0.4V富集80s后,以120mV/s扫描速度进行测定,结果表明:在1.0×10^-7-1.0×10^-4mol·L^-1浓度范围内,Bi^3+的溶出峰电流Ip与其呈良好的线性关系,检出限为6.0×10^-8mol·L^-1(信噪比RSN=3)。将该法应用于药物中铋含量的测定,回收率在99—102%,结果令人满意。%The anodic stripping volt - ampere characteristic of Bi^3 + on the carbon paste electrode was studied. The results show that there is a well defined stripping peak at -0.055V in the HCl solution( 1.0mol L^-1), and the electrode reaction is an adsorptioncontrolled process. In addition, we investigated the optimal condition of detecting Bi^3+ . The performance was carried at the scan rate of 120mV/s in the HCI solution ( 1.0mol L^-1 ) after accumulating tbr 80s at 0.4V. The results show the calibration curve was linear in the Bi^3 + concentration range of 1.0 × 10^-7 - 1.0×10 4 mol · L^-1 with a detection limit of 6.0 × 10 ^-8mol L 1 ( RsN = 3). When the proposed method was applied in determining the content of Bi^3 + in drugs, the results were satisfactory for that the recove was at 99% - 102%.