首页> 中文期刊> 《大连医科大学学报》 >罗氏 e601电化学发光仪剩余试剂的再利用评价

罗氏 e601电化学发光仪剩余试剂的再利用评价

         

摘要

Objective To evaluate the recycling value of remaining reagent by Roche e601 electrochemiluminescence ana-lyzer.Methods The precision, accuracy, comparison and correlation analysis were performed between original reagents and recycling reagents for HBsAb、HBcAb、HBeAg、and HBeAb.Results The intra batch precision results of level 1 were 3.43, 3.88,2.04,1.22 respectively,the intra batch precision results of level 2 were 1.22,2.40,1.09,1.66 respectively;the inter batch precision results of level 1 were 5.17,4.55,3.70,3.74 respectively;the inter batch precision results of level 2 were 4.99,3.93,4.16,6.91 respectively.The accuracy result was consistent with the EQA resuthe correlation coefficient of HBsAb was 0.999;the positive rate and negative coincidence rate(%)of HBcAb,HBeAg,HBeAb were 100%.Conclusion The remaining reagents can be combined and recycled by Roche e601, demonstrate good stability and ensure the quality of measurement.%目的:对罗氏e601电化学发光仪剩余试剂再利用进行评估。方法将HBsAb、HBcAb、HBeAg、HBeAb的混合剩余试剂,进行精密度、正确度验证并与新试剂进行比对及相关性分析。结果剩余混合试剂HBsAb、HB-cAb、HBeAg、HBeAb水平1批内精密度分别为3.43、3.88、2.04、1.22,水平2批内精密度分别为1.22、2.40、1.09、1.66;HBsAb、HBcAb、HBeAg、HBeAb水平1批间精密度分别为5.17、4.55、3.70、3.74;水平2批间精密度分别为4.99、3.93、4.16、6.91;正确度本室测定结果与室间质评回报结果均相符、HBsAb相关系数r为0.999;HBcAb、HBeAg、HBeAb两试剂盒的阳性符合率和阴性符合率均为100%。结论罗氏e601电化学发光仪剩余试剂可以混合使用,稳定性良好,能保证检测质量。