首页> 中文期刊> 《航空航天医学杂志》 >雅培ARCHITECT i2000 SR型全自动免疫发光仪测量3种肿瘤标记物的性能验证

雅培ARCHITECT i2000 SR型全自动免疫发光仪测量3种肿瘤标记物的性能验证

         

摘要

目的:通过对科室购入的雅培ARCHITECT i2000 SR型全自动免疫发光仪3种瘤标志物的性能验证,探讨全自动免疫发光仪合适有效的验证方案和实验方法。方法按CLSI 推荐的方法测定3种肿瘤标志物(CAl25Ⅱ,CA153,CAl99-XR)的精密度、准确度、性范围、生物参考区间,并进行验证。结果3项肿瘤标志物的方法学评价结果均符合本科室的质量要求。结论雅培ARCHITECT i2000化学发光分析仪检测系统性能满足临床和科室要求,评价方案科学适当。仪器应按要求保养,每年至少进行一次性能验证。设备商提供的数据必须经过科室严格的性能评价,以验证厂家声明,保证实验结果准确可靠。%Objective By validating the performance of three kinds tumor marks to explore appropriate and effective programs of verification and experimental methods for automatic immune luminescence instrument .Results Three tumor markers methodological evaluation results are in line with the quality of the undergraduate chamber .Conclusions The performance of Abbott ARCHITECT i 2000 chemiluminescent analyzer system meet the requirements of clinical and labo -ratory.The evaluating programs are scientific and appropriate .The instrument must be maintained according to the re-quirements and complete performance verification at least once a year.Data supplied by equipment must undergo rigorous performance evaluation of department to verify the declare of manufacturers so ensuring the accurate and reliable results .

著录项

相似文献

  • 中文文献
  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号