食管电极导管使用次数与食管调搏起搏电压阈值的关系

摘要

Objective To explore the relationship between the frequency of use of esophageal electrode catheter and transesophageal cardiac electrophysiology pacing threshold voltage. Methods A total of 123 patients who had esophageal pacing check during 2012 - 2013 in Daping Hospital were enrolled in this study. This study compared the pacing thresholds in groups A (having used for 1 - 2 times),B(having used for 3 - 4 times),C(having used for more than 4 times). Results The aver-age pacing threshold voltage was(13. 07 ± 2. 63)V in group A,(17. 15 ± 2. 60)V in group B,(21. 65 ± 1. 89)V in group C. The difference was significant(F = 147. 019,P < 0. 05);and group B was higher than group A,group C was higher than group A,B(P < 0. 05). Conclusion Esophageal electrode catheter will lead to the increase of pacing threshold due to passi-vation coating of surface of the electrode when it using for more than 3 - 4 times,which hinders its promotion and use in clinical practice.%目的:探讨食管电极导管的使用次数与食管调搏起搏电压阈值的关系。方法选取2012-2013年在重庆第三军医大学大坪医院行食管调搏检查的患者123例,比较采用新导管或曾使用过1~2次(A 组)、3~4次(B组)、4次以上(C 组)食管电极导管患者的起搏电压阈值。结果 A 组患者的起搏电压阈值为(13.07±2.63)V,B组患者的起搏电压阈值为(17.15±2.60)V,C 组患者的起搏电压阈值为(21.65±1.89)V,3组患者的起搏电压阈值间差异有统计学意义(F =147.019,P <0.05);其中 B 组起搏电压阈值高于 A 组,C 组起搏电压阈值高于 A、B 组,差异均有统计学意义(P <0.05)。结论食管电极导管重复使用3~4次以上时因电极表面钝化膜的形成导致起搏电压阈值增高,阻碍了该项技术在临床的推广与使用。