With the development of science and technology, medical equipments used in clinical diagnosis plays a very important role, is the STA-compact automatic coagulometer common faults are analyzed, and the fault reasons are discussed. Through practical work in STA-compact full automatic blood coagulation analyzer appears common fault cause analysis, combined with the given examples of troubleshooting for repair. The STA-compact full automatic blood coagulation instrument repair, find the essence of the problem, so in practical work on the service life of equipment, use efficiency is enhanced obviously. Through the introduction of STA-compact fully automatic blood coagulation analyzer system constitute, function and three kinds of common fault, aims to provide the operator of related reference information.%例举实践工作中STA-compact全自动血凝仪的系统构成、功能及常见3种故障,分析其故障现象,并对产生原因进行探讨.通过结合维修实例给出故障排除方案,总结维修经验为操作者提供相关参考信息.