【24h】

A new method for the quantitative analysis of gated SPECT polar scintigraphic maps

机译:门控SPECT极谱图定量分析的新方法

获取原文
获取原文并翻译 | 示例

摘要

We have developed a software, which allows to do non conventional percent quantitative analysis on scintigraphic polar map obtained from conventional processing of gated-SPECT acquisitions. Polar maps are 8 bit images of perfusion, motion, ejection fraction (EF) and thickening, of the heart The software is written in Matlab, analyses the whole polar map and four ROIs corresponding to the theoretical LAD, LCX, RCA territories (perfused by these arteries) and extra-ROIs region. An intensity segmentation is performed. The area corresponding to pixels lower and higher than a varying cut-off are calculated on the whole image and for each ROI. The software calculates an intensity-area histogram, which is the analogous of the Dose-Volume Histogram used in radiation therapy: in this case, the histogram has the meaning of a Perfusion- or a Motion-Volume histogram. Then, the software applies the Lyman-Wolbarst algorithm, to calculate the area equivalent histogram reduction (e.g. the perfused area in the hypothesis that all pixels are perfused at 100%.). The makes a direct comparison between two different polar maps by choice. The comparison between the numerical quantification of motion and perfusion maps, allows the physicians to get a clinical evaluation of the stunned myocardium.
机译:我们已经开发了一个软件,该软件可以对通过门控SPECT采集的常规处理获得的闪烁极谱图进行非常规百分比定量分析。极坐标图是心脏的灌注,运动,射血分数(EF)和增厚的8位图像。该软件用Matlab编写,分析了整个极坐标图和对应于理论LAD,LCX,RCA区域的四个ROI(这些动脉)和额外的投资回报率区域。进行强度分割。在整个图像上并针对每个ROI计算与低于和低于变化的截止点的像素对应的区域。该软件将计算强度区域直方图,该强度直方图类似于放射疗法中使用的剂量-体积直方图:在这种情况下,直方图具有“灌注”或“运动-体积”直方图的含义。然后,该软件应用Lyman-Wolbarst算法来计算面积等效直方图的缩减量(例如,假设所有像素均以100%进行灌注的假设中的灌注面积)。通过选择,可以在两个不同的极坐标图之间进行直接比较。运动图和灌注图的数字量化之间的比较,使医生能够对昏迷的心肌进行临床评估。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号