首页> 外文会议>Physics of Medical Imaging pt.2; Progress in Biomedical Optics and Imaging; vol.8 no.29; Proceedings of SPIE-The International Society for Optical Engineering; vol.6510 pt.2 >Dedicated dental volumetric and total body multislice computed tomography: a comparison of image quality and radiation dose
【24h】

Dedicated dental volumetric and total body multislice computed tomography: a comparison of image quality and radiation dose

机译:专用牙科体积和全身多层计算机断层扫描:图像质量和放射剂量的比较

获取原文
获取原文并翻译 | 示例

摘要

Aim of this work is to compare the performances of a Xoran Technologies i-CAT Cone Beam CT for dental applications with those of a standard total body multislice CT (Toshiba Aquilion 64 multislice) used for dental examinations. Image quality and doses to patients have been compared for the three main i-CAT protocols, the Toshiba standard protocol and a Toshiba modified protocol. Images of two phantoms have been acquired: a standard CT quality control phantom and an Alderson Rando~R anthropomorphic phantom. Image noise, Signal to Noise Ratio (SNR), Contrast to Noise Ratio (CNR) and geometric accuracy have been considered. Clinical image quality was assessed. Effective dose and doses to main head and neck organs were evaluated by means of thermo-luminescent dosimeters (TLD-100) placed in the anthropomorphic phantom. A Quality Index (QI), defined as the ratio of squared CNR to effective dose, has been evaluated. The evaluated effective doses range from 0.06 mSv (i-CAT 10 s protocol) to 2.37 mSv (Toshiba standard protocol). The Toshiba modified protocol (halved tube current, higher pitch value) imparts lower effective dose (0.99 mSv). The conventional CT device provides lower image noise and better SNR, but clinical effectiveness similar to that of dedicated dental CT (comparable CNR and clinical judgment). Consequently, QI values are much higher for this second CT scanner. No geometric distortion has been observed with both devices. As a conclusion, dental volumetric CT supplies adequate image quality to clinical purposes, at doses that are really lower than those imparted by a conventional CT device.
机译:这项工作的目的是比较Xoran Technologies i-CAT锥形束CT在牙科应用上的性能与用于牙科检查的标准全身多层CT(Toshiba Aquilion 64多层)的性能。对于三种主要的i-CAT协议,东芝标准协议和东芝改进协议,已经比较了患者的图像质量和剂量。已获取两个模型的图像:一个标准的CT质量控制模型和一个Alderson Rando〜R拟人模型。已经考虑了图像噪声,信噪比(SNR),对比度与噪声比(CNR)和几何精度。临床图像质量进行了评估。通过放置在拟人模型中的热辐射剂量计(TLD-100)来评估有效剂量和对主要头颈部器官的剂量。已经评估了质量指数(QI),即CNR与有效剂量的比值。评估的有效剂量范围为0.06 mSv(i-CAT 10 s方案)至2.37 mSv(Toshiba标准方案)。东芝改进的方案(减半的管电流,较高的螺距值)赋予较低的有效剂量(0.99 mSv)。传统的CT设备可提供较低的图像噪声和更好的SNR,但其临床效果类似于专用牙科CT(可比的CNR和临床判断)。因此,此第二台CT扫描仪的QI值要高得多。两种器件均未观察到几何变形。结论是,牙科容积CT以真正低于常规CT装置的剂量为临床目的提供了足够的图像质量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号