Simulation of countrate performance for a PET scanner with partial collimation

机译:部分准直的PET扫描仪的计数性能模拟

获取原文
获取原文并翻译 | 示例

摘要

We present a simulation study of the countrate performance of a PET scanner with partial collimation. In this study, partial collimation is achieved by removal of every other septum from the standard 2D septa set for the GE Advance PET scanner. System behavior is evaluated with a photon tracking simulation package (SimSET) and calibrated to measured data for 2D and fully-3D acquisition modes using the NEMA NU-2 countrate phantom. Results are evaluated in terms of true, scattered, and random coincidences and noise equivalent counts (NEC) regarding both counts per image plane and total counts as a function of activity. Our results show a good agreement between the measured and simulated count rates for the Advance PET scanner for the 2D (full collimation) and fully-3D (no collimation) acquisition modes, increasing our confidence in the predicted countrate results for the partial collimation mode. The latter results in a countrate performance intermediate between the 2D and fully-3D acquisition modes and yields a more favorable countrate performance for clinical activity levels.
机译:我们提出了部分准直的PET扫描仪计数性能的模拟研究。在这项研究中,通过从GE Advance PET扫描仪设置的标准2D隔垫中去除隔垫来实现部分准直。系统行为使用光子跟踪仿真程序包(SimSET)进行评估,并使用NEMA NU-2计数速率模型对2D和全3D采集模式下的测量数据进行校准。根据真实,分散和随机巧合以及噪声等效计数(NEC)来评估结果,这涉及每个图像平面的计数和总计数作为活动的函数。我们的结果表明,对于2D(完全准直)和完全3D(不准直)采集模式,Advance PET扫描仪的测量值与模拟计数率之间有很好的一致性,从而提高了我们对部分准直模式的预测计数结果的信心。后者导致2D采集和完全3D采集模式之间的计数率性能介于中间,并针对临床活动水平产生更有利的计数率性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号